Sánchez L, Regh M, Biesterfeld S, Chatelain R, Böcking A
Institute of Pathology, Aachen University of Technology, Federal Republic of Germany.
Anal Quant Cytol Histol. 1990 Aug;12(4):279-84.
The performance of a TV image analysis system combined with an automated microscope (the Leitz TAS plus) as a microdensitometer and morphometric device was investigated. There was a strict linear correlation (r greater than .99) between the physically defined optical transmission values and the resulting electronic signals from the plumbicon TV camera for the whole area displayed on the monitor. The shading of the optoelectronic system had a coefficient of variation (CV) of 1.42% for measurements in the center of the displayed area, but a CV of 3.55% for measurements over the whole monitor area. Densitometric measurements remained stable 15 minutes after putting the microscope lamp into operation (T less than 0.075%, remeasuring every two minutes). The geometric distortion, measured as different ferret diameters of ideally round latex particles, ranged from +/- 0.5% to +/- 1.0% deviation over the entire displayed area. These results indicate that densitometric and morphometric measurements with this equipment are sufficiently precise and reproducible when performed in the center of the area displayed on the monitor.
研究了将电视图像分析系统与自动显微镜(Leitz TAS plus)相结合作为微密度计和形态测量设备的性能。对于显示器上显示的整个区域,物理定义的光透射值与来自氧化铅电视摄像机的所得电子信号之间存在严格的线性相关性(r大于0.99)。对于显示区域中心的测量,光电系统的阴影变异系数(CV)为1.42%,但对于整个显示器区域的测量,CV为3.55%。在显微镜灯投入运行15分钟后(每两分钟重新测量一次),密度测量保持稳定(T小于0.075%)。以理想圆形乳胶颗粒的不同雪貂直径测量的几何畸变在整个显示区域的偏差范围为±0.5%至±1.0%。这些结果表明,当在显示器上显示区域的中心进行测量时,使用该设备进行密度测量和形态测量具有足够的精度和可重复性。