An indirect method to compare the reference centres for corneal measurements.

机构信息

Wenzhou Medical College, Wenzhou, China.

出版信息

Ophthalmic Physiol Opt. 2012 Mar;32(2):125-32. doi: 10.1111/j.1475-1313.2011.00880.x. Epub 2011 Nov 9.

Abstract

PURPOSE

Corneal measurements are commonly presented with respect to a specific reference centre, but the location of the reference centre on the corneal surface could vary from one diagnostic modality to another. This study aimed to develop a method for comparing reference centres used by corneal measurement systems.

METHODS

An indirect method was developed to compare reference centres by making use of the pupil centre and its offset from the reference centre. Reference centres in a Scheimpflug imaging system, the Pentacam HR, and a Placido-ring corneal topography system, the ATLAS Corneal Topography System, were compared for the right and left eyes of 30 subjects. The subjects all had similar pupil sizes when measured by the two systems. Differences and correlations of the pupil centre offsets between the two systems were statistically tested and compared by Bland-Altman analyses.

RESULTS

There were no significant differences in mean pupil offsets between the two systems for either the right or left eyes (p > 0.05). There were strong correlations of the pupil centre offsets between the two systems for each eye (right eye x-axis: r = 0.95, p < 0.0001; right eye y-axis: r = 0.98, p < 0.0001; left eye x-axis: r = 0.96, p < 0.0001, left eye y-axis: r = 0.93, p < 0.0001). Bland-Altman analyses revealed no significant differences in pupil centre offsets between the two systems.

CONCLUSIONS

The Pentacam HR system and the ATLAS system have very similar reference centres. Thus it is possible to directly analyze data from the Pentacam HR and other instruments using the corneal vertex or the pupil centre as the reference centres due to the similarity in the reference centre settings between the two systems.

摘要

目的

角膜测量通常是相对于特定的参考中心进行的,但参考中心在角膜表面的位置可能因诊断模态的不同而有所变化。本研究旨在开发一种比较角膜测量系统使用的参考中心的方法。

方法

通过利用瞳孔中心及其与参考中心的偏移,开发了一种间接比较参考中心的方法。比较了Scheimpflug 成像系统(Pentacam HR)和 Placido 环角膜地形图系统(ATLAS Corneal Topography System)的参考中心,共 30 名受试者的右眼和左眼。这两个系统测量的受试者瞳孔大小相似。通过 Bland-Altman 分析对两个系统的瞳孔中心偏移的差异和相关性进行了统计学检验和比较。

结果

两个系统的右眼和左眼的平均瞳孔偏移均无显著差异(p > 0.05)。两个系统的瞳孔中心偏移之间存在很强的相关性(右眼 x 轴:r = 0.95,p < 0.0001;右眼 y 轴:r = 0.98,p < 0.0001;左眼 x 轴:r = 0.96,p < 0.0001,左眼 y 轴:r = 0.93,p < 0.0001)。Bland-Altman 分析显示两个系统的瞳孔中心偏移没有显著差异。

结论

Pentacam HR 系统和 ATLAS 系统具有非常相似的参考中心。因此,由于两个系统的参考中心设置相似,可以直接分析 Pentacam HR 和其他仪器的数据,使用角膜顶点或瞳孔中心作为参考中心。