980nm 高功率二极管激光消融前列腺时 Twister 激光光纤的降解和组织消融能力。与标准侧射光纤的随机研究。

The Twister laser fiber degradation and tissue ablation capability during 980-nm high-power diode laser ablation of the prostate. A randomized study versus the standard side-firing fiber.

机构信息

Urology Department, Ain Shams University, Cairo, Egypt.

出版信息

Lasers Med Sci. 2012 Sep;27(5):959-63. doi: 10.1007/s10103-011-1017-8. Epub 2011 Nov 10.

Abstract

The objective of this work is to test the ablation capability and fiber degradation of the novel Twister fibers (TW), in both the large (LTW) and the standard (STW) sizes, against the standard side-firing (SF) fiber in a clinical setting during the treatment of BPH patients using the 980-nm high-power diode laser (HPDL). One hundred and twenty BPH patients treated with HPDL (Ceralase300, Biolitec AG, Jena, Germany) were randomized to receive treatment by one of the three fibers. Operative time corrected to tissue volume, laser treatment time, and laser energy were measured. Ablation rate was calculated as follows: the decrease of the prostate volume after 6 months/laser time. The fibers' resistance to degradation was defined by the laser energy needed to degrade the fiber completely. Preoperative prostate volume of 76 ± 38, 70 ± 39, and 88 ± 49 cc decreased by 49 ± 16, 51 ± 20, and 63 ± 16% for the SF, STW, and LTW fibers, respectively. This difference was highly significant when the LTW was compared to the other two fibers (p < 0.001). Prostate volume reduction post-operatively within each group as compared to the pre-operative volume was highly significant (p < 0.001). The ablation rate was highest in LTW, being 1.31 ± 0.59, 1.09 ± 0.51, and 1.54 ± 0.44 cc/min for the SF, STW, and LTW fibers, respectively. The LTW fiber resisted degradation more than the other fibers and the STW more than the SF fiber (p < 0.001). This study demonstrates the higher ablation efficiency and resistance to degradation of the LTW fiber as compared to the STW and SF fibers. The STW fiber has a similar ablation rate of the SF fiber but resists degradation better.

摘要

本研究旨在测试新型 Twister 光纤(TW)在临床治疗 BPH 患者时的消融能力和纤维降解情况,所用光纤包括大尺寸 Twister 光纤(LTW)和标准尺寸 Twister 光纤(STW),以及标准侧发光光纤(SF),激光为 980nm 高功率二极管激光(HPDL)。120 例 BPH 患者随机接受三种光纤之一的治疗。记录手术时间(校正至组织体积)、激光治疗时间和激光能量。消融率的计算方法为:前列腺体积 6 个月后的减少量除以激光时间。纤维的降解阻力定义为完全降解纤维所需的激光能量。术前前列腺体积为 76±38、70±39 和 88±49cc,分别减少了 49±16、51±20 和 63±16%,SF、STW 和 LTW 纤维的差异具有统计学意义(p<0.001)。与其他两种纤维相比,LTW 纤维的前列腺体积减少量显著更高(p<0.001)。与术前相比,术后每组的前列腺体积减少量均具有统计学意义(p<0.001)。LTW 纤维的消融率最高,为 1.31±0.59、1.09±0.51 和 1.54±0.44cc/min,SF、STW 和 LTW 纤维的消融率依次降低。LTW 纤维的降解阻力明显高于其他两种纤维,STW 纤维的降解阻力明显高于 SF 纤维(p<0.001)。本研究表明,与 STW 和 SF 纤维相比,LTW 纤维具有更高的消融效率和更强的抗降解能力。STW 纤维的消融率与 SF 纤维相似,但抗降解能力更强。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/fcb7/3414700/861ad59e0b9a/10103_2011_1017_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验