Wendt-Nordahl Gunnar, Huckele Stephanie, Honeck Patrick, Alken Peter, Knoll Thomas, Michel Maurice Stephan, Häcker Axel
Department of Urology, University Hospital Mannheim, Mannheim, Germany.
Eur Urol. 2007 Dec;52(6):1723-8. doi: 10.1016/j.eururo.2007.06.029. Epub 2007 Jun 26.
The wavelength 980 nm of a recently introduced diode laser system for treatment of benign prostatic enlargement offers a high simultaneous absorption in water and haemoglobin, and is postulated to combine high tissue ablative properties with a good haemostasis.
The Ceralas HPD150 diode laser system was evaluated in the well-established ex vivo model of the isolated blood-perfused porcine kidney to evaluate tissue ablation capacity and haemostatic properties at different generator settings. A histological examination of the ablated renal tissue followed. The results were compared with the reference standards transurethral resection of the prostate (TURP) and potassium-titanyl-phosphate (KTP) laser.
The diode laser displays a higher tissue ablation capacity, reaching 7.24+/-1.48 g after 10 min, compared with the KTP laser (3.99+/-0.48 g; p<0.05), whereas only 30s are needed to resect the tissue in the same surface area using TURP, resulting in 8.28+/-0.38 g of tissue removal. With a bleeding rate of 0.14+/-0.07 g/min, the diode laser offers haemostatic properties equivalent to the KTP laser (0.21+/-0.07 g/min) and a significantly reduced bleeding compared with TURP (20.14+/-2.03 g/min; p<0.05). The corresponding depths of the coagulation zones are 290.1+/-46.9 microm for the diode laser, 666.9+/-64.0 microm for the KTP laser (p<0.05), and 287.1+/-27.5 microm for TURP.
In the standardised ex vivo investigation, the 980-nm diode laser offers a higher tissue ablation capacity and similar haemostasis compared with the KTP laser. In comparison with TURP, both tissue ablation and bleeding are significantly reduced. The promising ex vivo results warrant further clinical investigation.
最近推出的用于治疗良性前列腺增生的二极管激光系统,其980纳米波长在水和血红蛋白中具有高同时吸收率,据推测它兼具高组织消融特性和良好的止血功能。
在成熟的离体血液灌注猪肾体外模型中评估Ceralas HPD150二极管激光系统,以评估不同发生器设置下的组织消融能力和止血特性。随后对消融的肾组织进行组织学检查。将结果与前列腺经尿道切除术(TURP)和磷酸钛钾(KTP)激光的参考标准进行比较。
与KTP激光(3.99±0.48克;p<0.05)相比,二极管激光显示出更高的组织消融能力,10分钟后达到7.24±1.48克,而使用TURP在相同表面积切除组织仅需30秒,切除组织量为8.28±0.38克。二极管激光的出血率为0.14±0.07克/分钟,其止血特性与KTP激光相当(0.21±0.07克/分钟),与TURP相比出血显著减少(20.14±2.03克/分钟;p<0.05)。二极管激光凝固区的相应深度为290.1±46.9微米,KTP激光为666.9±64.0微米(p<0.05),TURP为287.1±27.5微米。
在标准化的体外研究中,与KTP激光相比,980纳米二极管激光具有更高的组织消融能力和相似的止血效果。与TURP相比,组织消融和出血均显著减少。体外研究的良好结果值得进一步开展临床研究。