J Mech Behav Biomed Mater. 2012 Jan;5(1):238-46. doi: 10.1016/j.jmbbm.2011.08.027. Epub 2011 Sep 16.
Impaction bone grafting using morcellised allograft can successfully restore bone stock in revision surgery. However, concerns exist regarding supply of bone and transmission of infection. Bone-graft extenders, such as tricalcium phosphate (TCP) and hydroxyapatite (HA), are used to minimise the use of donor bone. However, concerns exist around a reported increased risk of femoral fracture during impaction bone grafting with a 1:1 mixture of TCP/HA and morcellised bone graft (MBG) during impaction grafting in human cadaveric femora. Using a sawbones model, it was evaluated whether there was increased femoral cortical strain with a HA:MBG mixture during impaction grafting compared to MBG impacted at the same and a greater force. Subsequently the subsidence behaviour of the different graft mixes was compared by using a loaded femoral stem in an endurance test. It was demonstrated that the femora with the MBG:HA graft had greater cortical hoop stresses but improved subsidence behaviour compared to a graft composed of pure MBG impacted at the same force.
使用切碎同种异体移植物进行嵌压植骨可以成功地在翻修手术中恢复骨量。然而,人们对骨的供应和感染的传播存在担忧。骨移植增强剂,如磷酸三钙 (TCP) 和羟基磷灰石 (HA),被用于尽量减少供骨的使用。然而,在人体尸体股骨的嵌压植骨中,使用 TCP/HA 和切碎骨移植物 (MBG) 的 1:1 混合物进行嵌压植骨时,据报道股骨骨折的风险增加,这引起了人们的关注。使用仿骨模型,评估了在相同和更大的力下进行嵌压植骨时,与 MBG 相比,HA:MBG 混合物是否会增加股骨皮质应变。随后,通过在耐力试验中使用带负荷的股骨柄比较了不同移植物混合物的下沉行为。结果表明,与在相同力下冲击的纯 MBG 移植物相比,MBG:HA 移植物的皮质环向应力更大,但沉降行为得到改善。