Laboratory of Analytical Chemistry, Department of Chemistry, Aristotle University of Thessaloniki, Thessaloniki, Greece.
Anal Bioanal Chem. 2012 May;403(4):1039-48. doi: 10.1007/s00216-011-5543-9. Epub 2011 Nov 20.
A new fluorescence-HPLC method was developed for the simultaneous determination of eight biogenic monoamines (histamine, methylamine, tyramine, ethylamine, propylamine, tryptamine, 2-phenylethylamine, isoamylamine) and two biogenic diamines (putrescine, cadaverine) in the presence of heptylamine as the internal standard. The amines were pre-column derivatized with naphthalene-2,3-dicarboxaldehyde in the presence of cyanide ion as the nucleophile. The effect of the derivatization reaction conditions on the reaction yield was investigated. The derivatives were separated on an Inertsil ODS-3 column (250 × 4mm i.d., 5 μm) using gradient elution and detected fluorimetrically at excitation and emission wavelengths of 424 and 494 nm, respectively. Limits of detection between 0.002 and 0.4 ng, injected on-column (10-μL loop), were achieved. The within- and between-day relative standard deviations ranged between 0.2-3.4% and 0.3-4.8%, respectively. The utility of the method in assaying biogenic mono- and diamine mixtures in Greek cheeses is demonstrated. Ultrasound-assisted liquid-liquid extraction was applied prior to derivatization.
建立了一种新的荧光高效液相色谱法,用于同时测定八种生物单胺(组氨酸、甲胺、酪胺、乙胺、丙胺、色胺、2-苯乙胺、异戊胺)和两种生物二胺(腐胺、尸胺)。内标物为正庚胺。在氰离子作为亲核试剂的存在下,用萘-2,3-二醛将胺进行柱前衍生化。考察了衍生化反应条件对反应收率的影响。衍生物在 Inertsil ODS-3 柱(250×4mm i.d.,5μm)上用梯度洗脱分离,在激发和发射波长分别为 424nm 和 494nm 处进行荧光检测。在 10μL 进样环上,检测限在 0.002-0.4ng 之间。日内和日间相对标准偏差分别在 0.2-3.4%和 0.3-4.8%之间。该方法在测定希腊奶酪中生物单胺和二胺混合物中的应用得到了验证。在衍生化前采用超声辅助液-液萃取。