Suppr超能文献

新型脚泵输尿管镜冲洗系统的系统评价。

Systematic evaluation of a novel foot-pump ureteroscopic irrigation system.

机构信息

Department of Urologic Surgery, University of Minnesota, Minneapolis, MN, USA.

出版信息

J Endourol. 2012 Feb;26(2):126-9. doi: 10.1089/end.2011.0248. Epub 2011 Dec 13.

Abstract

INTRODUCTION

To evaluate forces exerted on a stone with different ureteroscopic irrigation systems.

METHODS

A 3 mm steel simulated stone was welded to a 3F stone basket and inserted into the working channel of a 4.5F Wolf semirigid ureteroscope. The basket shaft was attached to a 50 g load cell. The urterescope was placed in a ureteral model (14F silicon tubing). Simulated blood (McCormick™ Red Dye, 1 dye drop/15 mL H(2)O) was dripped adjacent to the stone at 12 drops/min. Endoirrigation devices were attached to the ureterscope and irrigation was applied at a rate sufficient to maintain visualization of the stone. Force on the stone was measured with the following endoirrigation systems: Boston Scientific™ Single-Action-Pump System (SAP) hand-pump and NuVista Medical™ Flo-Assist(®) foot-pump.

RESULTS

No significant difference (p=0.19) in the number of pumps required to maintain a clear endoscopic field was found between the SAP (0.20±0.075/second) and Flo-Assist device (0.25±0.056/second). The pump duration of the Flo-Assist was found to be less (1.12±0.40 seconds) than the SAP (1.35±0.31 seconds), but not significantly different (p=0.24). The average maximum impulse was significantly lower (p=0.0002) for the SAP (8.34×10(-5) Ns) than the Flo-Assist (1.96×10(-3) Ns). Total maximum impulse (2.02×10(-5) Ns) for the SAP and total average impulse (5.51×10(-6) Ns) were found to be lower than the Flo-Assist device. The Flo-Assist had a total maximum impulse of 4.49×10(-4) Ns and total average impulse of 8.85×10(-6) Ns, however, these differences were not statistically significant.

CONCLUSION

The hand-pump (SAP) and foot-pump (Flo-Assist) irrigation devices tested require comparable amounts of pumps for similar durations of time to maintain endoscopic visualization. Overall, the SAP device exerts less average maximum force on the stone than the Flo-Assist device.

摘要

介绍

评估不同输尿管镜冲洗系统对结石施加的力。

方法

将一个 3 毫米的钢模拟结石焊接到一个 3F 结石篮上,并插入到 4.5F Wolf 半刚性输尿管镜的工作通道中。篮子轴连接到一个 50 克的称重传感器上。输尿管镜放置在输尿管模型(14F 硅管)中。模拟血液(McCormick™红色染料,每 15 毫升水中 1 滴)以 12 滴/分钟的速度滴在结石旁边。将内置冲洗装置连接到输尿管镜上,并以足以维持结石可视化的速度进行冲洗。使用以下内置冲洗系统测量对结石的作用力:波士顿科学™单动作泵系统(SAP)手动泵和 NuVista Medical™ Flo-Assist®(®)脚踏泵。

结果

在维持清晰的内镜视野所需的泵数方面,SAP(0.20±0.075/秒)和 Flo-Assist 装置(0.25±0.056/秒)之间没有显著差异(p=0.19)。发现 Flo-Assist 的泵持续时间(1.12±0.40 秒)比 SAP(1.35±0.31 秒)短,但差异不显著(p=0.24)。SAP 的平均最大脉冲显著较低(p=0.0002)(8.34×10(-5) Ns),而 Flo-Assist 的平均最大脉冲为 1.96×10(-3) Ns。SAP 的总最大脉冲(2.02×10(-5) Ns)和总平均脉冲(5.51×10(-6) Ns)均低于 Flo-Assist 装置。Flo-Assist 的总最大脉冲为 4.49×10(-4) Ns,总平均脉冲为 8.85×10(-6) Ns,但这些差异无统计学意义。

结论

测试的手动泵(SAP)和脚踏泵(Flo-Assist)冲洗装置在维持内镜可视化方面需要类似的泵数和持续时间。总体而言,SAP 装置对结石的平均最大力小于 Flo-Assist 装置。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验