Department of Imaging Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas 77030, USA.
Med Phys. 2011 Dec;38(12):6429-42. doi: 10.1118/1.3656040.
In this study, we used a small field high resolution detector in conjunction with a full field flat panel detector to implement and investigate the dual detector volume-of-interest (VOI) cone beam breast computed tomography (CBCT) technique on a bench-top system. The potential of using this technique to image small calcifications without increasing the overall dose to the breast was demonstrated. Significant reduction of scatter components in the high resolution projection image data of the VOI was also shown.
With the regular flat panel based CBCT technique, exposures were made at 80 kVp to generate an air kerma of 6 mGys at the isocenter. With the dual detector VOI CBCT technique, a high resolution small field CMOS detector was used to scan a cylindrical VOI (2.5 cm in diameter and height, 4.5 cm off-center) with collimated x-rays at four times of regular exposure level. A flat panel detector was used for full field scan with low x-ray exposures at half of the regular exposure level. The low exposure full field image data were used to fill in the truncated space in the VOI scan data and generate a complete projection image set. The Feldkamp-Davis-Kress (FDK) filtered backprojection algorithm was used to reconstruct high resolution images for the VOI. Two scanning techniques, one breast centered and the other VOI centered, were implemented and investigated. Paraffin cylinders with embedded thin aluminum (Al) wires were imaged and used in conjunction with optically stimulated luminescence (OSL) dose measurements to demonstrate the ability of this technique to image small calcifications without increasing the mean glandular dose (MGD).
Using exposures that produce an air kerma of 6 mGys at the isocenter, the regular CBCT technique was able to resolve the cross-sections of Al wires as thin as 254 μm in diameter in the phantom. For the specific VOI studied, by increasing the exposure level by a factor of 4 for the VOI scan and reducing the exposure level by a factor of 2 for the full filed scan, the dual-detector CBCT technique was able to resolve the cross-sections of Al wires as thin as 152 μm in diameter. The CNR evaluated for the entire Al wire cross-section was found to be improved from 5.5 in regular CBCT to 14.4 and 16.8 with the breast centered and VOI centered scanning techniques, respectively. Even inside VOI center, the VOI scan resulted in significant dose saving with the dose reduced by a factor of 1.6 at the VOI center. Dose saving outside the VOI was substantial with the dose reduced by a factor of 7.3 and 7.8 at the breast center for the breast centered and VOI centered scans, respectively, when compared to full field scan at the same exposure level. The differences between the two dual detector techniques in terms of dose saving and scatter reduction were small with VOI scan at 4× exposure level and full field scan at 0.5 × exposure level. The MGDs were only 94% of that from the regular CBCT scan.
For the specific VOI studied, the dual detector VOI CBCT technique has the potential to provide high quality images inside the VOI with MGD similar to or even lower than that of full field breast CBCT. It was also found that our results were compromised by the use of inadequate detectors for the VOI scan. An appropriately selected detector would better optimize the image quality improvement that can be achieved with the VOI CBCT technique.
在这项研究中,我们使用小视野高分辨率探测器与全视野平板探测器相结合,在台式系统上实现并研究双探测器感兴趣区(VOI)锥形束乳腺计算机断层摄影(CBCT)技术。证明了该技术在不增加乳房总剂量的情况下对小钙化进行成像的潜力。还显示了 VOI 高分辨率投影图像数据中散射分量的显著减少。
使用常规平板 CBCT 技术,在 80 kVp 下进行曝光,在等中心处产生 6 mGys 的空气比释动能。使用双探测器 VOI CBCT 技术,使用高分辨率小视野 CMOS 探测器以四倍于常规曝光水平的准直 X 射线扫描圆柱形 VOI(直径和高度为 2.5 厘米,距中心 4.5 厘米)。使用平板探测器以常规曝光水平的一半进行低 X 射线曝光的全场扫描。低曝光全场图像数据用于填充 VOI 扫描数据中的截断空间,并生成完整的投影图像集。使用 Feldkamp-Davis-Kress(FDK)滤波反投影算法对 VOI 进行高分辨率图像重建。实施并研究了两种扫描技术,一种是乳房中心,另一种是 VOI 中心。使用嵌入薄铝(Al)线的石蜡圆柱体进行成像,并结合光激励发光(OSL)剂量测量,证明该技术具有在不增加平均腺体剂量(MGD)的情况下对小钙化进行成像的能力。
使用在等中心处产生 6 mGys 空气比释动能的曝光,常规 CBCT 技术能够解析体模中直径为 254 μm 的 Al 线的横截面。对于研究的特定 VOI,通过将 VOI 扫描的曝光水平提高 4 倍,并将全场扫描的曝光水平降低 2 倍,双探测器 CBCT 技术能够解析直径为 152 μm 的 Al 线的横截面。对于整个 Al 线横截面,发现 CNR 从常规 CBCT 的 5.5 提高到 14.4 和 16.8,分别采用乳房中心和 VOI 中心扫描技术。即使在 VOI 中心内部,VOI 扫描也能显著节省剂量,在 VOI 中心处的剂量减少了 1.6 倍。与全场扫描相比,在相同的曝光水平下,在乳房中心处的 VOI 中心和 VOI 中心扫描的剂量减少了 7.3 和 7.8 倍,在 VOI 外部的剂量节省非常显著。在 4×曝光水平的 VOI 扫描和 0.5×曝光水平的全场扫描下,两种双探测器技术在剂量节省和散射减少方面的差异很小。MGD 仅为常规 CBCT 扫描的 94%。
对于研究的特定 VOI,双探测器 VOI CBCT 技术具有在 VOI 内提供高质量图像的潜力,MGD 与全视野乳腺 CBCT 相似,甚至更低。还发现,我们的结果受到用于 VOI 扫描的不适当探测器的限制。选择合适的探测器将更好地优化可以通过 VOI CBCT 技术实现的图像质量改进。