Suppr超能文献

采用平板探测器移位的锥形束乳腺 CT

Cone-beam breast computed tomography with a displaced flat panel detector array.

机构信息

Università di Napoli Federico II, Dipartimento di Scienze Fisiche, Napoli, Italy.

出版信息

Med Phys. 2012 May;39(5):2805-19. doi: 10.1118/1.4704641.

Abstract

PURPOSE

In cone-beam computed tomography (CBCT), and in particular in cone-beam breast computed tomography (CBBCT), an important issue is the reduction of the image artifacts produced by photon scatter and the reduction of patient dose. In this work, the authors propose to apply the detector displacement technique (also known as asymmetric detector or "extended view" geometry) to approach this goal. Potentially, this type of geometry, and the accompanying use of a beam collimator to mask the unirradiated half-object in each projection, permits some reduction of radiation dose with respect to conventional CBBCT and a sizeable reduction of the overall amount of scatter in the object, for a fixed contrast-to-noise ratio (CNR).

METHODS

The authors consider a scan configuration in which the projection data are acquired from an asymmetrically positioned detector that covers only one half of the scan field of view. Monte Carlo simulations and measurements, with their CBBCT laboratory scanner, were performed using PMMA phantoms of cylindrical (70-mm diameter) and hemiellipsoidal (140-mm diameter) shape simulating the average pendant breast, at 80 kVp. Image quality was evaluated in terms of contrast, noise, CNR, contrast-to-noise ratio per unit of dose (CNRD), and spatial resolution as width of line spread function for high contrast details.

RESULTS

Reconstructed images with the asymmetric detector technique deviate less than 1% from reconstruction with a conventional symmetric detector (detector view) and indicate a reduction of the cupping artifact in CT slices. The maximum scatter-to-primary ratio at the center of the phantom decreases by about 50% for both small and large diameter phantoms (e.g., from 0.75 in detector view to 0.40 in extended view geometry at the central axis of the 140-mm diameter PMMA phantom). Less cupping produces an increase of the CT number accuracy and an improved image detail contrast, but the associated increase of noise observed may produce a decrease of detail CNR. By simulating the energy deposited inside the phantoms, the authors evaluated a maximum 50% reduction of the absorbed dose at the expense of a decrease of CNR, for the half beam irradiation of the object performed with the displaced detector technique with respect to full beam irradiation. The decrease in CNR, and in absorbed dose as well, translates into a detail CNRD showing values comparable to or higher than the ones obtained for a conventional symmetric detector technique, attributed to the effect of decreased scatter in particular at the axis of the irradiated object. An estimate is provided (about 12%) for the average dose reduction possible in CBBCT at constant CNR for the average uncompressed breast (14 cm diameter, 50% glandularity), in case of minimum image overlapping in extended view.

CONCLUSIONS

Simulations and experiments show that CBCT reconstructions with the displaced detector technique and with a half beam collimator are less affected by scatter artifacts, which could lead to some decrease of the radiation dose to the irradiated object with respect to a conventional reconstruction. This dose reduction is associated with increase of noise, decrease of CNR, but equal or improved CNRD values. The use of a small area detector would allow also to reduce the apparatus cost and to improve the data transfer speed with a corresponding increment of frame rate.

摘要

目的

在锥形束计算机断层扫描(CBCT)中,特别是在锥形束乳腺计算机断层扫描(CBBCT)中,一个重要的问题是减少由光子散射产生的图像伪影和降低患者剂量。在这项工作中,作者提出应用探测器位移技术(也称为不对称探测器或“扩展视图”几何形状)来实现这一目标。这种类型的几何形状,以及使用光束准直器在每个投影中遮挡未照射的半物体,可以在固定对比噪声比(CNR)的情况下,相对于传统的 CBBCT 减少一些辐射剂量,并大大减少物体中的散射总量。

方法

作者考虑了一种扫描配置,其中从仅覆盖扫描视场的一半的不对称位置的探测器获取投影数据。使用 PMMA 圆柱体(70 毫米直径)和半椭圆形(140 毫米直径)形状的圆柱形(70 毫米直径)和半椭圆形(140 毫米直径)的 PMMA 体模,在 80 kVp 下进行蒙特卡罗模拟和测量,模拟平均悬垂乳房。使用对比度、噪声、CNR、每单位剂量的对比度噪声比(CNRD)和高对比度细节的线扩展函数宽度来评估图像质量。

结果

使用不对称探测器技术重建的图像与使用传统对称探测器(探测器视图)重建的图像相差小于 1%,并且 CT 切片中的杯状伪影减少。对于小直径和大直径的体模(例如,在直径为 140 毫米的 PMMA 体模的中心轴上,从探测器视图的 0.75 减少到扩展视图几何形状的 0.40),体模中心的最大散射与初级射线比减少约 50%。较少的杯状伪影会提高 CT 数的准确性和图像细节对比度,但观察到的噪声增加可能会降低细节 CNR。通过模拟体模内沉积的能量,作者评估了在使用位移探测器技术对物体进行半束照射时,相对于全束照射,吸收剂量最大可降低 50%,但 CNR 降低。CNR 的降低以及吸收剂量的降低转化为细节 CNRD,显示出与传统对称探测器技术获得的值相当或更高的值,这归因于散射的降低,特别是在照射物体的轴上。对于平均未压缩乳房(直径 14 厘米,50%腺体),在扩展视图中最小图像重叠的情况下,对于平均 CNR 恒定的 CBBCT,提供了可能的平均剂量降低的估计(约 12%)。

结论

模拟和实验表明,使用位移探测器技术和半束准直器的 CBCT 重建受散射伪影的影响较小,这可能导致与传统重建相比,照射物体的辐射剂量降低。这种剂量降低与噪声增加、CNR 降低有关,但 CNRD 值相等或提高。使用小面积探测器还可以降低设备成本,并通过相应增加帧率来提高数据传输速度。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验