The development of intracavitary ultrasonic applicators for hyperthermia: a design and experimental study.

作者信息

Diederich C J, Hynynen K

机构信息

Department of Radiation Oncology, Arizona Health Sciences Center, Tucson 85724.

出版信息

Med Phys. 1990 Jul-Aug;17(4):626-34. doi: 10.1118/1.596582.

Abstract

This study investigated the design concepts and development of a multielement intracavitary ultrasound applicator for use in hyperthermia. A necessary condition imposed on these applicators is that each transducer element be separately powered and produce collimated beams. This way, the power deposition within the target volume can be controlled by varying the power to each element. Theoretical computer simulations (acoustic and thermal) and bench experiments were used to determine the constraints on the transducer element size and the spacing between them. These have shown that the length of the cylindrical segments (or subsections of) must be greater than approximately 10 lambda for proper collimation and that the spacing between them must be less than approximately 1.5 mm for uniform heating. With these design principles in mind, applicators were constructed using sections of cylindrical transducers (wall-thickness resonance). These were surrounded by temperature-controlled circulating water which was enclosed by a latex membrane. This allowed for acoustic coupling and additional control over the depth of the maximum temperature from the cavity wall. This depth could be varied between the cavity surface and up to 1.5 cm for circulating water temperatures between 5 and 42 degrees C, respectively. These applicators were tested in vivo and were able to induce controlled transrectal heating, at depths of 2-3 cm, in the canine rectum and prostate gland.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验