第二代双能 CT 虚拟平扫:衰减值的可靠性。

Virtual non-contrast in second-generation, dual-energy computed tomography: reliability of attenuation values.

机构信息

Department of Radiology, Medical University Vienna/Vienna General Hospital, Vienna, Austria.

出版信息

Eur J Radiol. 2012 Mar;81(3):e398-405. doi: 10.1016/j.ejrad.2011.12.011. Epub 2012 Jan 10.

Abstract

PURPOSE

To evaluate the reliability of attenuation values in virtual non-contrast images (VNC) reconstructed from contrast-enhanced, dual-energy scans performed on a second-generation dual-energy CT scanner, compared to single-energy, non-contrast images (TNC).

MATERIALS AND METHODS

Sixteen phantoms containing a mixture of contrast agent and water at different attenuations (0-1400 HU) were investigated on a Definition Flash-CT scanner using a single-energy scan at 120 kV and a DE-CT protocol (100 kV/SN140 kV). For clinical assessment, 86 patients who received a dual-phase CT, containing an unenhanced single-energy scan at 120 kV and a contrast enhanced (110 ml Iomeron 400 mg/ml; 4 ml/s) DE-CT (100 kV/SN140 kV) in an arterial (n=43) or a venous phase, were retrospectively analyzed. Mean attenuation was measured within regions of interest of the phantoms and in different tissue types of the patients within the corresponding VNC and TNC images. Paired t-tests and Pearson correlation were used for statistical analysis.

RESULTS

For all phantoms, mean attenuation in VNC was 5.3±18.4 HU, with respect to water. In 86 patients overall, 2637 regions were measured in TNC and VNC images, with a mean difference between TNC and VNC of -3.6±8.3 HU. In 91.5% (n=2412) of all cases, absolute differences between TNC and VNC were under 15HU, and, in 75.3% (n=1986), differences were under 10 HU.

CONCLUSIONS

Second-generation dual-energy CT based VNC images provide attenuation values close to those of TNC. To avoid possible outliers multiple measurements are recommended especially for measurements in the spleen, the mesenteric fat, and the aorta.

摘要

目的

评估在第二代双能 CT 扫描仪上进行对比增强双能扫描重建的虚拟非对比图像(VNC)中的衰减值的可靠性,与单能非对比图像(TNC)相比。

材料与方法

在 Definition Flash-CT 扫描仪上使用 120kV 单能扫描和 DE-CT 方案(100kV/SN140kV)对包含不同衰减值(0-1400HU)的造影剂和水混合物的 16 个体模进行了研究。为了临床评估,对 86 名接受双期 CT 的患者进行了回顾性分析,其中包含未增强的 120kV 单能扫描和对比增强(110ml Iomeron 400mg/ml;4ml/s)的 DE-CT(100kV/SN140kV)在动脉期(n=43)或静脉期。在 VNC 和 TNC 图像中,在体模的感兴趣区域和患者的不同组织类型内测量平均衰减。采用配对 t 检验和 Pearson 相关分析进行统计学分析。

结果

对于所有体模,VNC 中的平均衰减值为 5.3±18.4HU,相对于水。在 86 名患者中,在 TNC 和 VNC 图像中测量了 2637 个区域,TNC 和 VNC 之间的平均差值为-3.6±8.3HU。在 91.5%(n=2412)的所有病例中,TNC 和 VNC 之间的绝对差值小于 15HU,在 75.3%(n=1986)的病例中,差值小于 10HU。

结论

第二代双能 CT 基于 VNC 图像提供的衰减值与 TNC 非常接近。为了避免可能的异常值,建议进行多次测量,特别是在脾脏、肠系膜脂肪和主动脉中进行测量。