Department of Oral and Maxilloficial Radiology, University of Texas Health Science Centre at San Antonio, TX, USA.
Dentomaxillofac Radiol. 2012 Mar;41(3):248-53. doi: 10.1259/dmfr/80899839. Epub 2012 Jan 12.
Metal in a patient's mouth has been shown to cause artefacts that can interfere with the diagnostic quality of cone beam CT. Recently, a manufacturer has made an algorithm and software available which reduces metal streak artefact (Picasso Master 3D® machine; Vatech, Hwaseong, Republic of Korea).
The purpose of this investigation was to determine whether or not the metal artefact reduction algorithm was effective and enhanced the contrast-to-noise ratio.
A phantom was constructed incorporating three metallic beads and three epoxy resin-based bone substitutes to simulate bone next to metal. The phantom was placed in the centre of the field of view and at the periphery. 10 data sets were acquired at 50-90 kVp. The images obtained were analysed using a public domain software ImageJ (NIH Image, Bethesda, MD). Profile lines were used to evaluate grey level changes and area histograms were used to evaluate contrast. The contrast-to-noise ratio was calculated.
The metal artefact reduction option reduced grey value variation and increased the contrast-to-noise ratio. The grey value varied least when the phantom was in the middle of the volume and the metal artefact reduction was activated. The image quality improved as the peak kilovoltage increased.
Better images of a phantom were obtained when the metal artefact reduction algorithm was used.
患者口腔中的金属已被证明会导致伪影,从而干扰锥形束 CT 的诊断质量。最近,制造商提供了一种算法和软件,可减少金属条纹伪影(Picasso Master 3D®机器;韩国华城的 Vatech)。
本研究旨在确定金属伪影减少算法是否有效并提高了对比噪声比。
构建了一个包含三个金属珠和三个基于环氧树脂的骨替代物的体模,以模拟金属旁边的骨头。将体模放置在视野的中心和外围。在 50-90 kVp 下采集了 10 组数据。使用公共域软件 ImageJ(NIH Image,贝塞斯达,MD)分析获得的图像。使用轮廓线评估灰度变化,使用面积直方图评估对比度。计算对比噪声比。
金属伪影减少选项减少了灰度值变化并提高了对比噪声比。当体模位于体积中心且激活金属伪影减少功能时,灰度值变化最小。随着峰值千伏的增加,图像质量得到改善。
当使用金属伪影减少算法时,可获得更好的体模图像。