Analysis of post-exposure density growth in radiochromic film with respect to the radiation dose.

机构信息

Department of Radiation Oncology and Medical Physics, Graduate School of Medicine, Sapporo Medical University, Japan.

出版信息

J Radiat Res. 2012;53(2):301-5. doi: 10.1269/jrr.11091. Epub 2012 Jan 14.

Abstract

The post-exposure density growth (PEDG) is one of the characteristics of radiochromic film (RCF). In film dosimetry using RCF and a flatbed scanner, pixel values read out from the RCF are converted to dose (hereafter, film dose) by using a calibration curve. The aim of this study is to analyze the relationship between the pixel value read out from the RCF and the PEDG, and that between the film dose converted from the RCF and the PEDG. The film (GAFCHROMIC EBT) was irradiated with 10-MV X-rays in an ascending 11-dose-step arrangement. The pixel values of the irradiated EBT film were measured at arbitrary hours using an Epson flatbed scanner. In this study, the reference time was 24 h after irradiation, and all dose conversions from the pixel values read out from the EBT film were made using a calibration curve for 24 h after irradiation. For delivered doses of 33 and 348 cGy, the measured pixel values at 0.1 and 16 h after irradiation represented ranges of -9.6% to -0.7% and -3.9% to -0.3%, respectively, of the reference value. The relative changes between the pixel values read out from the EBT film at each elapsed time and that at the reference time decreased with increasing delivered dose. However, the difference range for all the film doses had a width of approximately -10% of the reference value at elapsed times from 0.1 to 16 h, and it showed no dependence on the delivered dose.

摘要

曝光后密度增长(PEDG)是光致变色胶片(RCF)的特征之一。在使用 RCF 和平板扫描仪进行胶片剂量测定时,通过使用校准曲线将从 RCF 读取的像素值转换为剂量(以下简称胶片剂量)。本研究旨在分析从 RCF 读取的像素值与 PEDG 之间的关系,以及从 RCF 转换的胶片剂量与 PEDG 之间的关系。胶片(GAFCHROMIC EBT)用 10-MV X 射线在递增的 11 剂量步长布置下进行照射。用爱普生平板扫描仪在任意小时测量照射的 EBT 胶片的像素值。在本研究中,参考时间为照射后 24 小时,并且使用照射后 24 小时的校准曲线对从 EBT 胶片读取的像素值进行了所有剂量转换。对于 33 和 348 cGy 的传递剂量,在照射后 0.1 和 16 小时测量的像素值分别代表参考值的-9.6%至-0.7%和-3.9%至-0.3%的范围。从 EBT 胶片读取的像素值在每个经过时间与参考时间之间的相对变化随着传递剂量的增加而减小。然而,在 0.1 至 16 小时的经过时间内,所有胶片剂量的差异范围都在参考值的-10%左右,并且与传递剂量无关。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验