Magn Reson Med. 2012 Sep;68(3):932-46. doi: 10.1002/mrm.23309. Epub 2012 Jan 13.
MR acoustic radiation force imaging (ARFI) is an elegant adjunct to MR-guided high intensity focused ultrasound for treatment planning and optimization, permitting in situ assessment of the focusing and targeting quality. The thermal effect of high intensity focused ultrasound pulses associated with ARFI measurements is recommended to be monitored on line, in particular when the beam crosses highly absorbent structures or interfaces (e.g., bones or air-filled cavities). A dedicated MR sequence is proposed here, derived from a segmented gradient echo-echo planar imaging kernel by adding a bipolar motion encoding gradient with interleaved alternating polarities. Temporal resolution was reduced to 2.1 s, with in-plane spatial resolution of 1 mm. MR-ARFI measurements were executed during controlled animal breathing, with trans-costal successively steered foci, to investigate the spatial modulation of the focus intensity and the targeting offset. ARFI-induced tissue displacement measurements enabled the accurate localization, in vivo, of the high intensity focused ultrasound focal point in sheep liver, with simultaneous monitoring of the temperature elevation. ARFI-based precalibration of the focal point position was immediately followed by trans-costal MR-guided high intensity focused ultrasound ablation, monitored with a conventional proton resonance frequency shift MR thermometry sequence. The latter MR thermometry sequence had spatial resolution and geometrical distortion identical with the ARFI maps, hence no coregistration was required.
磁共振声辐射力成像(ARFI)是磁共振引导高强度聚焦超声治疗计划和优化的一个很好的辅助手段,允许在原位评估聚焦和靶向质量。建议在线监测高强度聚焦超声脉冲的热效应与 ARFI 测量相关联,特别是当光束穿过高吸收结构或界面(例如,骨骼或充满空气的腔)时。这里提出了一种专用的磁共振序列,通过在分段梯度回波-回波平面成像内核中添加具有交替极性的双极运动编码梯度来派生。时间分辨率降低到 2.1 秒,平面内空间分辨率为 1 毫米。MR-ARFI 测量在受控动物呼吸期间执行,采用经肋部依次引导的焦点,以研究焦点强度和靶向偏移的空间调制。ARFI 诱导的组织位移测量能够在体内准确定位绵羊肝脏中的高强度聚焦超声焦点,同时监测温度升高。基于 ARFI 的焦点位置预校准立即由经肋部磁共振引导高强度聚焦超声消融进行,使用常规质子共振频率偏移磁共振测温序列进行监测。后者的磁共振测温序列具有与 ARFI 图谱相同的空间分辨率和几何变形,因此不需要配准。