Xu Feng, Helfen Lukas, Baumbach Tilo, Suhonen Heikki
Institute for Synchrotron Radiation, Karlsruhe Institute of Technology, Germany.
Opt Express. 2012 Jan 16;20(2):794-806. doi: 10.1364/OE.20.000794.
In computed tomography (CT), projection images of the sample are acquired over an angular range between 180 to 360 degrees around a rotation axis. A special case of CT is that of limited-angle CT, where some of the rotation angles are inaccessible, leading to artefacts in the reconstrucion because of missing information. The case of flat samples is considered, where the projection angles that are close to the sample surface are either i) completely unavailable or ii) very noisy due to the limited transmission at these angles. Computed laminography (CL) is an imaging technique especially suited for flat samples. CL is a generalization of CT that uses a rotation axis tilted by less than 90 degrees with respect to the incident beam. Thus CL avoids using projections from angles closest to the sample surface. We make a quantitative comparison of the imaging artefacts between CL and limited-angle CT for the case of a parallel-beam geometry. Both experimental and simulated images are used to characterize the effect of the artefacts on the resolution and visible image features. The results indicate that CL has an advantage over CT in cases when the missing angular range is a significant portion of the total angular range. In the case when the quality of the projections is limited by noise, CT allows a better tradeoff between the noise level and the missing angular range.
在计算机断层扫描(CT)中,围绕旋转轴在180度至360度的角度范围内采集样品的投影图像。CT的一种特殊情况是有限角度CT,其中一些旋转角度无法获取,由于信息缺失会导致重建中出现伪影。本文考虑了扁平样品的情况,其中靠近样品表面的投影角度要么i)完全不可用,要么ii)由于这些角度的传输受限而噪声很大。计算机分层摄影(CL)是一种特别适用于扁平样品的成像技术。CL是CT的一种推广,它使用相对于入射光束倾斜小于90度的旋转轴。因此,CL避免使用最接近样品表面角度的投影。对于平行光束几何形状的情况,我们对CL和有限角度CT之间的成像伪影进行了定量比较。实验图像和模拟图像都用于表征伪影对分辨率和可见图像特征的影响。结果表明,当缺失的角度范围占总角度范围的很大一部分时,CL比CT具有优势。在投影质量受噪声限制的情况下,CT在噪声水平和缺失角度范围之间能实现更好的权衡。