To determine the flow characteristics of the 3 different models of the Ex-PRESS miniature glaucoma device in a controlled laboratory study.
The 3 different Ex-PRESS models (P-50, R-50, and P-200; Optonol Ltd; now Alcon Lab) were tested using a gravity-driven flow test. Three samples of each of the 3 Ex-PRESS models were subjected to a constant gravitational force of fluid at 5 different pressure levels (5 to 25 mm Hg). Four measurements per sample were taken at each pressure level. The main outcome measure was flow rate (Q) (µL/min). Resistance (R) was calculated by dividing pressure (P) by the measured flow (Q).
The flow rate was primarily pressure dependent. The P-200 model (internal diameter 200 µm) showed a statistically significant higher flow rate and lower resistance compared with both the P-50 and R-50 models (internal diameter 50 µm) (P<0.0001). The P-50 and R-50 models demonstrated similar flow rates (P=0.08) despite their difference in tube length (2.64 vs. 2.94 mm).
The 3 models of the Ex-PRESS mini shunt behaved in vitro as simple flow resistors by creating a relatively constant resistance to flow. Tube diameter was the only parameter with significant impact on flow and resistance. All models demonstrated flow rates per unit of pressure much higher than the outflow facility of a healthy human eye.
在一项对照实验室研究中,确定 3 种不同型号的 Ex-PRESS 微型青光眼引流装置的流动特性。
使用重力驱动流量测试对 3 种不同的 Ex-PRESS 模型(P-50、R-50 和 P-200;Optonol Ltd;现为 Alcon Lab)进行了测试。对 3 种 Ex-PRESS 模型中的每一种的 3 个样本,在 5 个不同的压力水平(5 至 25 毫米汞柱)下施加恒定的流体重力。在每个压力水平下,对每个样本进行 4 次测量。主要观察指标是流量(Q)(µL/min)。阻力(R)通过压力(P)除以测量流量(Q)计算得出。
流量主要依赖于压力。与 P-50 和 R-50 模型(内径 50 µm)相比,P-200 模型(内径 200 µm)的流量更高,阻力更低,差异具有统计学意义(P<0.0001)。尽管 P-50 和 R-50 模型的管长(2.64 与 2.94 毫米)不同,但它们的流量相似(P=0.08)。