Department of Radiology, Kinki University, Ohnohigashi, Osakasayama, Osaka, Japan.
Ann Nucl Med. 2012 Apr;26(3):241-7. doi: 10.1007/s12149-011-0567-y. Epub 2012 Jan 26.
The aim of this study was to elucidate the regional differences between brain perfusion single photon emission computed tomography (SPECT) images reconstructed with a uniform attenuation correction using Chang's method (AC-Chang) and a non-uniform attenuation correction with CT using SPECT/CT (AC-CT).
SPECT images of a phantom with and without head holder were obtained, and reconstructed images of AC-Chang and AC-CT were compared. Twenty-eight consecutive patients with brain disease examined by SPECT/CT brain perfusion imaging were selected, and images were reconstructed with AC-Chang and AC-CT. The AC-Chang and AC-CT reconstructed images were then compared by voxel-based analysis using three-dimensional stereotactic surface projections.
Counts in the frontal area of the AC-Chang phantom image with head holder were higher than those in the posterior area. Counts in the frontal area of the AC-Chang clinical images were significantly higher than those in the AC-CT images, while the counts in the margin of the frontal lobe and posterior margin of the parietal, occipital cortices and cerebellum of the AC-Chang images were significantly lower. Relative frontal perfusion was 5.0% higher and relative cerebellar perfusion was 4.6% lower in the AC-Chang images relative to the AC-CT images, on average.
We demonstrated the frontal dominant hyper-perfusion and parieto-occipital and cerebellar hypo-perfusion in brain SPECT images reconstructed with AC-Chang compared to those reconstructed with AC-CT. We suggest that to obtain an accurate attenuation-corrected brain perfusion SPECT image, attenuation correction by Chang's method is inadequate.
本研究旨在阐明使用 Chang 法(AC-Chang)进行均匀衰减校正的脑灌注单光子发射计算机断层扫描(SPECT)图像与使用 SPECT/CT 的非均匀衰减校正(AC-CT)之间的区域差异。
获得有和没有头架的体模的 SPECT 图像,并比较 AC-Chang 和 AC-CT 的重建图像。选择 28 例连续的脑疾病患者进行 SPECT/CT 脑灌注成像检查,并用 AC-Chang 和 AC-CT 重建图像。然后通过三维立体表面投影的体素基分析比较 AC-Chang 和 AC-CT 重建图像。
有头架的 AC-Chang 体模图像的额区计数高于后区。AC-Chang 临床图像的额区计数明显高于 AC-CT 图像,而额叶边缘和顶、枕叶皮质及小脑后缘的 AC-Chang 图像计数明显较低。AC-Chang 图像相对于 AC-CT 图像的相对额叶灌注平均高 5.0%,相对小脑灌注低 4.6%。
与使用 AC-CT 重建的图像相比,我们在使用 AC-Chang 重建的脑 SPECT 图像中显示出额部优势性高灌注和顶枕叶及小脑低灌注。我们建议,为了获得准确的衰减校正脑灌注 SPECT 图像,Chang 法的衰减校正不足。