Department of Radiology, Hua Dong Hospital of Fudan University, Shanghai 200040, China.
Eur J Radiol. 2012 Oct;81(10):2597-601. doi: 10.1016/j.ejrad.2011.12.041. Epub 2012 Feb 1.
To compare image quality and visualization of normal structures and lesions in brain computed tomography (CT) with adaptive statistical iterative reconstruction (ASIR) and filtered back projection (FBP) reconstruction techniques in different X-ray tube current-time products.
In this IRB-approved prospective study, forty patients (nineteen men, twenty-one women; mean age 69.5 ± 11.2 years) received brain scan at different tube current-time products (300 and 200 mAs) in 64-section multi-detector CT (GE, Discovery CT750 HD). Images were reconstructed with FBP and four levels of ASIR-FBP blending. Two radiologists (please note that our hospital is renowned for its geriatric medicine department, and these two radiologists are more experienced in chronic cerebral vascular disease than in neoplastic disease, so this research did not contain cerebral tumors but as a discussion) assessed all the reconstructed images for visibility of normal structures, lesion conspicuity, image contrast and diagnostic confidence in a blinded and randomized manner. Volume CT dose index (CTDI(vol)) and dose-length product (DLP) were recorded. All the data were analyzed by using SPSS 13.0 statistical analysis software.
There was no statistically significant difference between the image qualities at 200 mAs with 50% ASIR blending technique and 300 mAs with FBP technique (p>.05). While between the image qualities at 200 mAs with FBP and 300 mAs with FBP technique a statistically significant difference (p<.05) was found.
ASIR provided same image quality and diagnostic ability in brain imaging with greater than 30% dose reduction compared with FBP reconstruction technique.
比较自适应统计迭代重建(ASIR)和滤波反投影(FBP)重建技术在不同 X 射线管电流-时间产品下脑 CT 图像质量和正常结构及病变的可视化效果。
本研究为经伦理委员会批准的前瞻性研究,共纳入 40 例患者(男 19 例,女 21 例;平均年龄 69.5±11.2 岁),在 64 层多排 CT(GE,Discovery CT750 HD)上分别采用 300 mAs 和 200 mAs 管电流-时间产品进行颅脑扫描。图像分别采用 FBP 和 4 种 ASIR-FBP 混合水平进行重建。由 2 名放射科医生(请注意,我们医院以老年医学科而闻名,这 2 名放射科医生在慢性脑血管疾病方面的经验多于肿瘤疾病,因此本研究不包含脑肿瘤,但作为讨论内容)以盲法和随机的方式评估所有重建图像的正常结构可见度、病变显示、图像对比度和诊断信心。记录容积 CT 剂量指数(CTDI(vol))和剂量长度乘积(DLP)。所有数据均采用 SPSS 13.0 统计分析软件进行分析。
200 mAs 时 50%ASIR 混合技术与 300 mAs 时 FBP 技术的图像质量无统计学差异(p>.05)。而 200 mAs 时 FBP 与 300 mAs 时 FBP 技术的图像质量有统计学差异(p<.05)。
与 FBP 重建技术相比,ASIR 可在减少 30%以上剂量的情况下提供相同的脑成像图像质量和诊断能力。