Holroyd John R, Edyvean Sue
1 Centre for Radiation, Chemical and Environmental Hazards (CRCE), Public Health England, Chilton , Didcot, Oxfordshire , UK.
Br J Radiol. 2018 May;91(1085):20170834. doi: 10.1259/bjr.20170834. Epub 2018 Feb 27.
To review doses to patients undergoing cervical spine CT examinations in the UK.
A data collection form was developed and distributed to medical physicists and radiographers via e-mail distribution lists. The form requested details of CT scanners, exposure protocols and patient dose index information.
Data were received for 73 scanners. It was seen that 97% of scanners used automatic exposure control, and 60% of scanners used an iterative reconstruction technique for cervical spine examinations. The majority of scans were taken at 120 kV. The average patient dose indicators in terms of CT dose index (CTDI) ranged from 3.5 to 39.7 mGy (mean value 16.7 mGy), and for the DLP, ranged from 87 to 1030 mGy cm (mean value 379 mGy cm) as quoted for the standard 32 cm phantom.
The rounded third quartile value of the mean dose distributions from this study were a CT dose index (CTDI) of 20 mGy and a dose-length product of 440 mGy cm as quoted for a 32 cm body phantom. These are significantly higher than those in the 2011 Public Health England CT dose survey when adjusted for phantom size. It is suggested that the existing national diagnostic reference levels for cervical spine CT should be amended, both with the new values and also to quote according to the 32 cm phantom. Advances in knowledge: Proposed new national diagnostic reference levels are presented for cervical spine CT examinations.
回顾英国颈椎CT检查患者所接受的剂量。
设计了一份数据收集表,并通过电子邮件分发列表分发给医学物理师和放射技师。该表格要求提供CT扫描仪、曝光协议和患者剂量指数信息的详细情况。
收到了73台扫描仪的数据。可以看出,97%的扫描仪使用自动曝光控制,60%的扫描仪在颈椎检查中使用迭代重建技术。大多数扫描在120 kV下进行。以CT剂量指数(CTDI)衡量的平均患者剂量指标范围为3.5至39.7 mGy(平均值16.7 mGy),对于剂量长度乘积(DLP),以标准32 cm体模为例,范围为87至1030 mGy cm(平均值379 mGy cm)。
本研究中平均剂量分布的四分之三分位数约为:对于32 cm体模,CT剂量指数(CTDI)为20 mGy,剂量长度乘积为440 mGy cm。经体模尺寸调整后,这些值显著高于2011年英国公共卫生部门CT剂量调查中的值。建议修订现有的颈椎CT国家诊断参考水平,包括新值,并按32 cm体模进行标注。知识进展:提出了颈椎CT检查的国家诊断参考水平新建议。