Placement of esophageal stethoscope by acoustic criteria does not consistently yield an optimal location for the monitoring of core temperature.

作者信息

Freund P R, Brengelmann G L

机构信息

Department of Anesthesiology, University of Washington, Seattle.

出版信息

J Clin Monit. 1990 Oct;6(4):266-70. doi: 10.1007/BF02842485.

Abstract

The esophageal stethoscope has evolved into a device for both acoustic and core temperature monitoring. To test whether routine placement according to acoustic criteria results in placement of the core temperature sensor in the region of contiguity between the esophagus and the heart, we determined the depth of placement electrocardiographically. All patients were undergoing nonthoracic elective operations requiring general anesthesia and tracheal intubation. First, we established that different observers selected the same esophageal depth within +/- 1 cm electrocardiographically, using the criterion of a symmetric biphasic P wave of maximal amplitude (7 patients). Then, in 30 more patients, we compared routine acoustic placements with the depths of the maximal-amplitude biphasic P wave. Stethoscopes placed according to acoustic criteria were within +/- 3 cm of P-wave depths in 15 of 30 patients. In the remaining patients, measured discrepancies ranged up to 13.5 cm. We conclude that the prevailing stethoscope design, with a thermistor at the tip, below the acoustic window, does not ensure placement of the thermistor within the optimal region for monitoring of core temperature. A modification in design that would take advantage of the reliability of electrocardiographic positioning is suggested.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验