We report the results of 150 capsular bag implantations of the IOGEL PC-12 hydrogel lens. Mean follow-up was seven months. Following capsulorhexis and phacoemulsification, a Faulkner folder was used to insert the lens through a 3.5 mm to 4.0 mm scleral tunnel incision. Initially, a number of surgical complications were encountered. Because of increasing experience and modified instrumentation and technique, these did not occur in the later cases. Visual results were good, all eyes gaining a best case visual acuity of 20/40 or better and 97% achieving 20/25 or better. Morphological results were satisfactory. Generally the lenses remained centered and at a distance from the iris. Retention of viscoelastic substance or debris between the lens and the posterior capsule, occasionally observed at the start of the series, has been avoided by retrolental aspiration. Persistent pigment dispersion, which was observed when the lens had been implanted in the sulcus, was not seen. Tolerance of the lens material was generally satisfactory. There were six cases (4%) of fibrinoid uveitis, which is a relatively high incidence. From our experience and results we conclude that this implantation procedure allowed controlled insertion and placement of the IOGEL lens, provided that adequate instrumentation and technique was used; visual performance was comparable to that of poly(methyl methacrylate) lenses; capsular bag fixation furnished satisfactory morphological results, provided an adequately shaped capsulorhexis was performed.
我们报告了150例IOGEL PC - 12水凝胶晶状体囊袋内植入术的结果。平均随访时间为7个月。在完成连续环形撕囊和超声乳化术后,使用福克纳夹通过3.5毫米至4.0毫米的巩膜隧道切口植入晶状体。最初,遇到了一些手术并发症。由于经验的增加以及器械和技术的改进,后期病例未再出现这些并发症。视觉效果良好,所有术眼的最佳视力均达到20/40或更好,97%的术眼达到20/25或更好。形态学结果令人满意。通常,晶状体保持居中且与虹膜有一定距离。通过晶状体后抽吸,避免了在该系列手术开始时偶尔观察到的晶状体与后囊膜之间的粘弹性物质或碎屑残留。当晶状体植入睫状沟时观察到的持续性色素播散现象未再出现。晶状体材料的耐受性总体令人满意。有6例(4%)发生了纤维蛋白样葡萄膜炎,这一发生率相对较高。根据我们的经验和结果,我们得出结论,只要使用了适当的器械和技术,这种植入方法能够实现IOGEL晶状体的可控植入和放置;视觉性能与聚甲基丙烯酸甲酯晶状体相当;囊袋固定提供了令人满意的形态学结果,前提是进行了形状合适的连续环形撕囊。