Stability of voxel values from cone-beam computed tomography for dental use in evaluating bone mineral content.

机构信息

Oral and Maxillofacial Radiology, Tokyo Medical and Dental University, Tokyo, Japan.

出版信息

Clin Oral Implants Res. 2013 May;24(5):543-8. doi: 10.1111/j.1600-0501.2012.02420.x. Epub 2012 Feb 10.

Abstract

OBJECTIVES

The objective of this study was to investigate the stability of voxel values from cone-beam CT (CBCT) using a flat panel detector in changing surrounding circumstances mimicking clinical situations.

MATERIALS AND METHODS

Seven hydroxyapatite (HA) rods (0, 40, 80, 120, 160, 199, and 239 mg Ca/cm(3) ) were set in a 16 cm diameter water phantom and scanned with CBCT (FineCube) and also with multislice CT (MSCT) used as reference. The voxel values from CBCT were measured by the obtained images with OsiriX software (http://www.osirix-viewer.com), and compared with the CT numbers from MSCT. We examined the correlation between the voxel values and the HA contents, the variance of voxel values, and the influence of changing surrounding circumstances on the voxel values.

RESULTS

The voxel values from CBCT were different from the CT numbers from MSCT, but there was a linear Pearson correlation between the voxel values and the HA contents. The voxel values were greatly influenced when the samples were scanned surrounded in air. However, the voxel values were not affected significantly in the existence of materials simulating oral hard tissues and/or a metal core post.

CONCLUSIONS

There was a linear correlation between the voxel values of CBCT and the contents of HA rod samples. Our study indicated that it might be possible to evaluate bone mineral content(BMC) from the voxel values of CBCT for dental implant treatment.

摘要

目的

本研究旨在通过平板探测器锥形束 CT(CBCT),在模拟临床情况的环境变化下,研究体素值的稳定性。

材料与方法

将 7 根羟磷灰石(HA)棒(0、40、80、120、160、199 和 239mgCa/cm(3))置于 16cm 直径水模中,分别使用 CBCT(FineCube)和多层 CT(MSCT)进行扫描作为参考。使用 OsiriX 软件(http://www.osirix-viewer.com)对 CBCT 获得的图像中的体素值进行测量,并与 MSCT 的 CT 值进行比较。我们检查了体素值与 HA 含量、体素值的变化、以及周围环境变化对体素值的影响之间的相关性。

结果

CBCT 的体素值与 MSCT 的 CT 值不同,但体素值与 HA 含量之间存在线性 Pearson 相关性。当样本在空气中扫描时,体素值受到很大影响。然而,在存在模拟口腔硬组织和/或金属核心柱的材料的情况下,体素值没有受到明显影响。

结论

CBCT 的体素值与 HA 棒样本的含量之间存在线性相关性。我们的研究表明,通过 CBCT 的体素值评估骨矿物质含量(BMC)可能用于牙科种植治疗。