用于密度估计的牙科锥形束 CT 灰度值的可变性。

Variability of dental cone beam CT grey values for density estimations.

机构信息

Department of Oral Health Sciences, University of Leuven, Leuven, Belgium.

出版信息

Br J Radiol. 2013 Jan;86(1021):20120135. doi: 10.1259/bjr.20120135.

Abstract

OBJECTIVE

The aim of this study was to investigate the use of dental cone beam CT (CBCT) grey values for density estimations by calculating the correlation with multislice CT (MSCT) values and the grey value error after recalibration.

METHODS

A polymethyl methacrylate (PMMA) phantom was developed containing inserts of different density: air, PMMA, hydroxyapatite (HA) 50 mg cm(-3), HA 100, HA 200 and aluminium. The phantom was scanned on 13 CBCT devices and 1 MSCT device. Correlation between CBCT grey values and CT numbers was calculated, and the average error of the CBCT values was estimated in the medium-density range after recalibration.

RESULTS

Pearson correlation coefficients ranged between 0.7014 and 0.9996 in the full-density range and between 0.5620 and 0.9991 in the medium-density range. The average error of CBCT voxel values in the medium-density range was between 35 and 1562.

CONCLUSION

Even though most CBCT devices showed a good overall correlation with CT numbers, large errors can be seen when using the grey values in a quantitative way. Although it could be possible to obtain pseudo-Hounsfield units from certain CBCTs, alternative methods of assessing bone tissue should be further investigated.

ADVANCES IN KNOWLEDGE

The suitability of dental CBCT for density estimations was assessed, involving a large number of devices and protocols. The possibility for grey value calibration was thoroughly investigated.

摘要

目的

本研究旨在通过计算与多层 CT(MSCT)值的相关性以及重新校准后的灰度值误差,来研究牙用锥形束 CT(CBCT)灰度值在密度估计中的应用。

方法

研制了一种包含不同密度嵌体的聚甲基丙烯酸甲酯(PMMA)体模:空气、PMMA、羟磷灰石(HA)50mg·cm(-3)、HA 100、HA 200 和铝。对 13 台 CBCT 设备和 1 台 MSCT 设备进行了扫描。计算了 CBCT 灰度值与 CT 值之间的相关性,并在重新校准后估计了中密度范围内的 CBCT 值的平均误差。

结果

在全密度范围内,Pearson 相关系数范围为 0.7014 至 0.9996,在中密度范围内,Pearson 相关系数范围为 0.5620 至 0.9991。中密度范围内 CBCT 体素值的平均误差在 35 至 1562 之间。

结论

尽管大多数 CBCT 设备与 CT 数值之间表现出良好的总体相关性,但在定量使用灰度值时会出现较大误差。虽然某些 CBCT 可能可以获得伪亨氏单位,但应进一步研究替代的骨组织评估方法。

知识进展

评估了牙用 CBCT 进行密度估计的适用性,涉及大量设备和方案。彻底研究了灰度值校准的可能性。

相似文献

1
Variability of dental cone beam CT grey values for density estimations.
Br J Radiol. 2013 Jan;86(1021):20120135. doi: 10.1259/bjr.20120135.
2
Bone quality evaluation at dental implant site using multislice CT, micro-CT, and cone beam CT.
Clin Oral Implants Res. 2015;26(1):e1-7. doi: 10.1111/clr.12315. Epub 2013 Dec 11.
3
Development and applicability of a quality control phantom for dental cone-beam CT.
J Appl Clin Med Phys. 2011 Nov 15;12(4):3478. doi: 10.1120/jacmp.v12i4.3478.
4
Artifacts in orthodontic bracket systems in cone-beam computed tomography and multislice computed tomography.
J Orofac Orthop. 2015 Mar;76(2):152-60, 162-3. doi: 10.1007/s00056-014-0278-9.
5
Stability of voxel values from cone-beam computed tomography for dental use in evaluating bone mineral content.
Clin Oral Implants Res. 2013 May;24(5):543-8. doi: 10.1111/j.1600-0501.2012.02420.x. Epub 2012 Feb 10.
6
Deriving Hounsfield units using grey levels in cone beam computed tomography.
Dentomaxillofac Radiol. 2010 Sep;39(6):323-35. doi: 10.1259/dmfr/19603304.
7
Influence of cone beam CT scanning parameters on grey value measurements at an implant site.
Dentomaxillofac Radiol. 2013;42(3):79884780. doi: 10.1259/dmfr/79884780. Epub 2012 Aug 29.
8
Reliability of voxel values from cone-beam computed tomography for dental use in evaluating bone mineral density.
Clin Oral Implants Res. 2010 May;21(5):558-62. doi: 10.1111/j.1600-0501.2009.01896.x.
9
Deriving Hounsfield units using grey levels in cone beam CT: a clinical application.
Dentomaxillofac Radiol. 2012 Sep;41(6):500-8. doi: 10.1259/dmfr/31640433. Epub 2012 Jun 29.
10
Investigation of dental cone-beam CT pixel data and a modified method for conversion to Hounsfield unit (HU).
Dentomaxillofac Radiol. 2018 Feb;47(2):20170321. doi: 10.1259/dmfr.20170321. Epub 2017 Nov 6.

引用本文的文献

2
Accuracy of the Hounsfield Unit Values Measured by Implant Planning Software.
Dent J (Basel). 2024 Dec 17;12(12):413. doi: 10.3390/dj12120413.
3
Automated bone property analysis using corrected in vivo dental cone-beam CT data of human wrists.
Sci Rep. 2024 Dec 16;14(1):30466. doi: 10.1038/s41598-024-75222-6.
7
Comparison of 96-kV and 120-kV cone-beam CT for the assessment of cochlear implants.
BMC Med Imaging. 2024 Jun 13;24(1):145. doi: 10.1186/s12880-024-01322-4.
9
Assessment of CBCT gray value in different regions-of-interest and fields-of-view compared to Hounsfield unit.
Dentomaxillofac Radiol. 2023 Nov;52(8):20230187. doi: 10.1259/dmfr.20230187. Epub 2023 Oct 24.

本文引用的文献

1
The detection of periapical pathosis using periapical radiography and cone beam computed tomography - part 1: pre-operative status.
Int Endod J. 2012 Aug;45(8):702-10. doi: 10.1111/j.1365-2591.2011.01989.x. Epub 2011 Dec 21.
2
Assessment of bone density in the posterior maxilla based on Hounsfield units to enhance the initial stability of implants.
Clin Implant Dent Relat Res. 2012 May;14 Suppl 1:e183-7. doi: 10.1111/j.1708-8208.2011.00423.x. Epub 2011 Dec 16.
3
Quantification of metal artifacts on cone beam computed tomography images.
Clin Oral Implants Res. 2013 Aug;24 Suppl A100:94-9. doi: 10.1111/j.1600-0501.2011.02382.x. Epub 2011 Dec 15.
4
Measuring bone density in healing periapical lesions by using cone beam computed tomography: a clinical investigation.
J Endod. 2012 Jan;38(1):28-31. doi: 10.1016/j.joen.2011.09.032. Epub 2011 Nov 13.
6
Development and applicability of a quality control phantom for dental cone-beam CT.
J Appl Clin Med Phys. 2011 Nov 15;12(4):3478. doi: 10.1120/jacmp.v12i4.3478.
9
Qualitative and quantitative evaluation of bony structures based on DICOM dataset.
J Oral Maxillofac Surg. 2011 Nov;69(11):2763-70. doi: 10.1016/j.joms.2011.02.076. Epub 2011 Jun 23.
10
Artefacts in CBCT: a review.
Dentomaxillofac Radiol. 2011 Jul;40(5):265-73. doi: 10.1259/dmfr/30642039.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验