Pabisch Silvia, Feichtenschlager Bernhard, Kickelbick Guido, Peterlik Herwig
University of Vienna, Faculty of Physics, 1090 Vienna, Austria.
Chem Phys Lett. 2012 Jan 10;521(C):91-97. doi: 10.1016/j.cplett.2011.11.049.
The aim of this work is a systematic comparison of size characterisation methods for two completely different model systems of oxide nanoparticles, i.e. amorphous spherical silica and anisotropic facet-shaped crystalline zirconia. Size and/or size distribution were determined in a wide range from 5 to 70 nm using small-angle X-ray scattering (SAXS), dynamic light scattering (DLS), nitrogen sorption (BET), X-ray diffraction (XRD) and transmission electron microscopy (TEM). A nearly perfect coincidence was observed only for SAXS and TEM for both types of particles. For zirconia nanoparticles considerable differences between different measurement methods were observed.
这项工作的目的是对两种完全不同的氧化物纳米颗粒模型系统,即非晶态球形二氧化硅和各向异性的晶面形氧化锆的尺寸表征方法进行系统比较。使用小角X射线散射(SAXS)、动态光散射(DLS)、氮吸附(BET)、X射线衍射(XRD)和透射电子显微镜(TEM)在5至70纳米的宽范围内测定尺寸和/或尺寸分布。对于这两种类型的颗粒,仅SAXS和TEM观察到几乎完全一致的结果。对于氧化锆纳米颗粒,观察到不同测量方法之间存在相当大的差异。