Ash Stephen R, Dhamija Rajiv, Zaroura Mohamad Y, Hentschel Dirk M
HemoCleanse, Inc. and Indiana University Health Arnett, Lafayette, Indiana 47904, USA.
Semin Dial. 2012 Jul;25(4):474-81. doi: 10.1111/j.1525-139X.2011.01037.x. Epub 2012 Feb 23.
The StenTec™ Gauge provides a method to determine the static intra-access pressure ratio (P(Ia Ratio) ) within a fistula or graft. The StenTec Gauge estimates the peak systolic pressure within the fistula or graft by measurement of the distance that the blood-air interface progresses into the tubing of the fistula needle, after the needle is inserted "dry" into the fistula or graft and before the cap is removed from the tubing. The peak systolic pressure is graphically compared with the systolic arterial blood pressure of the patient, to determine P(Ia Ratio) . For best accuracy, the StenTec Gauge should be chosen that best matches the internal volume of the fistula needle tubing (2.3-2.6 ml for 12-inch tubing and 3.6 ml for 16-inch tubing) and the approximate elevation of the city in which it is used (0-1000, 1000-3000, and 3000-6000 feet above sea level). In this article, we explain the rationale for this method of surveillance, evidence for accuracy of the StenTec Gauge, and the correlation of changes in the P(Ia Ratio) to the development of stenosis in a fistula or graft.
StenTec™测量仪提供了一种测定动静脉内瘘或移植物静态腔内压力比(P(Ia比率))的方法。StenTec测量仪通过测量在“无血”状态下将针头插入动静脉内瘘或移植物且尚未取下针管盖帽时,血液 - 空气界面在瘘针管内推进的距离,来估算动静脉内瘘或移植物内的收缩压峰值。将收缩压峰值与患者的动脉收缩压进行图形化比较,以确定P(Ia比率)。为获得最佳准确性,应选择与瘘针管内部容积(12英寸针管为2.3 - 2.6毫升,16英寸针管为3.6毫升)以及使用地所在城市的大致海拔高度(海平面以上0 - 1000英尺、1000 - 3000英尺和3000 - 6000英尺)最匹配的StenTec测量仪。在本文中,我们解释了这种监测方法的原理、StenTec测量仪准确性的证据,以及P(Ia比率)变化与动静脉内瘘或移植物狭窄发展的相关性。