Corrêa E L, Vivolo V, Potiens M P A
Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN), Avenida Professor Lineu Prestes 2242, Cidade Universitária, São Paulo, SP, Brazil.
Appl Radiat Isot. 2012 Jul;70(7):1396-9. doi: 10.1016/j.apradiso.2012.01.012. Epub 2012 Feb 15.
This study presents the results of the establishment of a quality control program developed and applied for the X-ray system of the Calibration Laboratory of IPEN. The X-ray standard beams, mammography level, using molybdenum and aluminum as additional filtration were established after the application of this quality control and the spectrometry of these qualities was made. The reference ionization chamber has traceability to the PTB. The radiation qualities RQR-M, RQA-M, RQN-M and RQB-M, following the recommendations of the IEC 61267 and the IAEA TRS 457 were established.