OBJECTIVES

To evaluate the influence of different biopsy forcipes on the deflection, irrigant flow, and optical characteristics of flexible ureterorenoscopes and to assess tissue quality for histopathologic evaluation in an ex vivo setting.

MATERIALS AND METHODS

The following five different biopsy forcipes were compared: Olympus (FB-56D-1; diameter 5Fr.), R. Wolf (829.601; 3Fr.), Karl Storz Medical (11275ZE; 3Fr.), Boston Scientific (Piranha; 505-160; 3 Fr.), and Cook BIGopsy (115CM; 2.4 Fr.). The devices were tested in 3 different ureterorenoscopes: Storz 11278 VU (Flex-X(2)), Storz 11278 V (Flex-X(C)), and Wolf Cobra (7326071/-6). Tissue samples were obtained from porcine upper urinary tracts.

RESULTS

Baseline irrigation flow rates with empty channels were significantly higher in the Wolf Cobra than in Storz ureterorenoscopes (30.5 vs. 23 and 21 ml/min). The BIGopsy forceps allowed for higher flow rates in both Storz ureterorenoscopes (2.2 and 1.3, respectively) when compared to the other devices (0.5 and 0.6 ml/min). The Storz and Wolf biopsy forcipes resulted in the highest impairment of the deflection angle. In all 3 ureterorenoscopes, flow rates and deflection angle were least impaired by the BIGopsy. However, BIGopsy compromised the field of view (20 % reduction vs. 12 % by others). The largest sample of renal pelvis and ureter biopsies was obtained with BIGopsy and Storz(®) forcipes, respectively. The extent of artifacts and denuded urothelium were comparable in all samples.

CONCLUSIONS

The various biopsy devices showed different impacts on irrigation flow, deflection, and field of view. The Cook BIGopsy best retains irrigation flow in single-channel flexible ureterorenoscopes and deflection. However, a smaller field of view may complicate handling and tissue acquisition.

摘要

目的

评估不同活检钳对软性输尿管肾镜的偏斜、冲洗液流动和光学特性的影响,并在离体环境下评估组织质量以进行组织病理学评估。

材料与方法

比较了以下 5 种不同的活检钳:Olympus(FB-56D-1;直径 5Fr.)、R. Wolf(829.601;3Fr.)、Karl Storz Medical(11275ZE;3Fr.)、Boston Scientific(Piranha;505-160;3Fr.)和 Cook BIGopsy(115CM;2.4Fr.)。将这些器械在 3 种不同的输尿管肾镜中进行测试:Storz 11278VU(Flex-X(2))、Storz 11278V(Flex-X(C))和 Wolf Cobra(7326071/-6)。组织样本取自猪的上尿路。

结果

空通道下的基础冲洗流量在 Wolf Cobra 输尿管肾镜中明显高于 Storz 输尿管肾镜(30.5 比 23 和 21ml/min)。与其他器械相比,BIGopsy 活检钳在 Storz 输尿管肾镜中可实现更高的流量(分别为 2.2 和 1.3ml/min)(0.5 和 0.6ml/min)。Storz 和 Wolf 活检钳导致偏斜角度的损害最大。在所有 3 种输尿管肾镜中,BIGopsy 对流量和偏斜角度的损害最小。然而,BIGopsy 会影响视野(比其他器械减少 20%)。最大的肾盂和输尿管活检样本分别用 BIGopsy 和 Storz(®)活检钳获得。所有样本的伪影和裸露的尿路上皮程度相当。

结论

各种活检器械对冲洗流量、偏斜和视野有不同的影响。Cook BIGopsy 在单通道软性输尿管肾镜中保持最佳的冲洗流量和偏斜度。然而,较小的视野可能会使操作和组织获取复杂化。