Department of Urology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
Ann Clin Biochem. 2012 May;49(Pt 3):273-6. doi: 10.1258/acb.2011.011046. Epub 2012 Mar 27.
Elevated alphafetoprotein (AFP) concentrations may result from a variety of clinical conditions, but their role as an important tumour marker has been well established. There may be differences in AFP values due to laboratories using different methods, even though most methods have been calibrated with the same international standard (WHO IS 72/225). Therefore it is important to know the analytical performance of the various methods in relation to the analytical requirements for AFP measurement.
Annually, from January 2005 to July 2010, the results were analysed from the 65-75 laboratories that took part in the AFP survey of the External Quality Assurance programme of the Foundation Quality Control Medical Laboratories (the SKML/Binding Analysis) in the Netherlands.
The Elecsys/Modular (36%) and the Immulite 2000/2500 (29%) are the methods used most. The methods show, on average, up to 15% positive and 12% negative bias, compared with the all-laboratory trimmed mean. Of the laboratories using the Immulite or the Elecsys/Modular analyser, over 70% show sufficient analytical performance to meet the Fraser criterion for method imprecision. Of the laboratories using a different method, over 50% do not meet this criterion.
AFP immunoassays suffer from method bias, even though all methods have been calibrated with the same international standard. Some of the methods used show insufficient performance.
甲胎蛋白(AFP)浓度升高可能由多种临床情况引起,但它作为重要肿瘤标志物的作用已得到充分证实。由于不同实验室使用的方法不同,可能会存在 AFP 值的差异,即使大多数方法都已经用相同的国际标准(世卫组织 72/225)进行了校准。因此,了解各种方法在 AFP 测量的分析要求方面的分析性能非常重要。
从 2005 年 1 月至 2010 年 7 月,每年对参加荷兰基金会质量控制医学实验室(SKML/结合分析)外部质量保证计划 AFP 调查的 65-75 个实验室的结果进行分析。
使用最广泛的方法是 Elecsys/Modular(36%)和 Immulite 2000/2500(29%)。与所有实验室修剪平均值相比,这些方法平均存在高达 15%的阳性和 12%的阴性偏倚。在使用 Immulite 或 Elecsys/Modular 分析仪的实验室中,超过 70%的实验室具有足够的分析性能,符合 Fraser 方法不精密度标准。在使用不同方法的实验室中,超过 50%的实验室不符合这一标准。
尽管所有方法都已用相同的国际标准校准,但 AFP 免疫测定仍存在方法偏倚。一些使用的方法显示出性能不足。