Department of Urology, Princess Alexandra Hospital, Ipswich Rd, Brisbane, QLD, 4103, Australia.
World J Urol. 2013 Jun;31(3):591-5. doi: 10.1007/s00345-012-0859-4. Epub 2012 Mar 29.
To evaluate the clinical outcomes of penile prosthesis implantation for the treatment for erectile dysfunction (ED) over 3 decades in a centre of excellence.
A total of 955 penile prostheses were implanted between June 1981 and June 2010. The mean age of the men was 53.2 (28-80) years, and the mean follow-up was 76 (12-355) months. A total of 771 men had primary implants. The most common implant was Ultrex cylinder (54 %), and the main cause of ED was organic (32 %).
Primary implants showed higher rate of intra-operative complications than revision surgery (3.5 vs. 0.1 %) (p < 0.05). Prosthesis infection occurred in 0.8 % and equal incidence between diabetic and pelvic trauma patients. The average time to prosthetic revision was 102 (30-210) months. Kaplan-Meier estimates of overall penile prosthesis survival at 5 and 10 years were around 90.8 and 85.0 %. The most common mechanical failures were fluid loss (75 %). The majority of men were satisfied with the surgical outcomes, and 90 % of men would undergo penile prosthesis implant again.
Penile prosthesis surgery is a safe and durable treatment option for male ED. Strict adherence to antimicrobial prophylaxis and surgical practice is paramount to ensure low complication rates and high patient satisfaction rate.
在卓越中心评估 30 多年来阴茎假体植入治疗勃起功能障碍(ED)的临床疗效。
1981 年 6 月至 2010 年 6 月期间共植入 955 件阴茎假体。男性的平均年龄为 53.2(28-80)岁,平均随访时间为 76(12-355)个月。771 名男性接受了初次植入。最常见的植入物是 Ultrex 圆柱体(54%),ED 的主要病因是器质性(32%)。
初次植入的术中并发症发生率高于翻修手术(3.5%比 0.1%)(p<0.05)。感染发生率为 0.8%,糖尿病和骨盆创伤患者的发生率相等。假体翻修的平均时间为 102(30-210)个月。Kaplan-Meier 估计的 5 年和 10 年阴茎假体总生存率分别约为 90.8%和 85.0%。最常见的机械故障是液体流失(75%)。大多数男性对手术结果满意,90%的男性会再次接受阴茎假体植入。
阴茎假体手术是男性 ED 的一种安全且持久的治疗选择。严格遵循抗菌预防和手术实践至关重要,以确保低并发症发生率和高患者满意度。