To evaluate the agreement of blood pressure measurements and hypertension scores obtained by use of 3 indirect arterial blood pressure measurement devices in hospitalized dogs. Design-Diagnostic test evaluation.
29 client-owned dogs.
5 to 7 consecutive blood pressure readings were obtained from each dog on each of 3 occasions with a Doppler ultrasonic flow detector, a standard oscillometric device (STO), and a high-definition oscillometric device (HDO).
When the individual sets of 5 to 7 readings were evaluated, the coefficient of variation for systolic arterial blood pressure (SAP) exceeded 20% for 0% (Doppler), 11 % (STO), and 28% (HDO) of the sets of readings. After readings that exceeded a 20% coefficient of variation were discarded, repeatability was within 25 (Doppler), 37 (STO), and 39 (HDO) mm Hg for SAP. Correlation of mean values among the devices was between 0.47 and 0.63. Compared with Doppler readings, STO underestimated and HDO overestimated SAP. Limits of agreement between mean readings of any 2 devices were wide. With the hypertension scale used to score SAP, the intraclass correlation of scores was 0.48. Linear-weighted inter-rater reliability between scores was 0.40 (Doppler vs STO), 0.38 (Doppler vs HDO), and 0.29 (STO vs HDO).
Results of this study suggested that no meaningful clinical comparison can be made between blood pressure readings obtained from the same dog with different indirect blood pressure measurement devices.
评估在住院犬中使用3种间接动脉血压测量设备所获得的血压测量值与高血压评分的一致性。设计-诊断试验评估。
29只客户拥有的犬。
分别使用多普勒超声血流探测器、标准示波测量设备(STO)和高清示波测量设备(HDO),在3个不同时间点对每只犬连续进行5至7次血压读数测量。
当对每组5至7次读数进行评估时,收缩期动脉血压(SAP)的变异系数在多普勒测量组中为0%,在STO测量组中为11%,在HDO测量组中为28%的读数组超过20%。在丢弃变异系数超过20%的读数后,SAP的重复性在多普勒测量组中为25 mmHg以内,在STO测量组中为37 mmHg以内,在HDO测量组中为39 mmHg以内。各设备之间平均值的相关性在0.47至0.63之间。与多普勒读数相比,STO低估了SAP,而HDO高估了SAP。任意两种设备平均读数之间的一致性界限较宽。使用高血压量表对SAP进行评分时,评分的组内相关性为0.48。评分之间的线性加权评分者间信度在多普勒与STO之间为0.40,在多普勒与HDO之间为0.38,在STO与HDO之间为0.29。
本研究结果表明,使用不同间接血压测量设备从同一只犬获得的血压读数之间无法进行有意义的临床比较。