Seliškar Alenka, Zrimšek Petra, Sredenšek Jerneja, Petrič Aleksandra D
Clinic for Small Animal Medicine and Surgery, Veterinary Faculty, University of Ljubljana, Ljubljana, Slovenia.
Vet Anaesth Analg. 2013 Jan;40(1):21-7. doi: 10.1111/j.1467-2995.2012.00774.x. Epub 2012 Sep 24.
To use the American College of Veterinary Internal Medicine (ACVIM) validation criteria to evaluate the performance of high definition oscillometric (HDO) and Doppler blood pressure measurement techniques against invasive blood pressure measurements in anaesthetized dogs.
Prospective clinical study.
Twenty client-owned dogs.
Invasive blood pressure was measured using a catheter inserted into a pedal artery and an electronic transducer. The sites of cuff placement for the HDO measurements were the mid antebrachium or the proximal tail and, for the Doppler technique, the distal tibia. Agreement between invasive and non-invasive blood pressure measurements was estimated by the Bland-Altman method.
Only 10% and 34% of Doppler measurements were within 10 and 20 mmHg of invasive blood pressure values, respectively. The Doppler device failed to meet the ACVIM validation criteria for blood pressure measurement devices. The best agreement between HDO and invasive blood pressure measurement technique was observed for mean arterial blood pressure (MAP); 67% and 95% of readings were within 10 and 20 mmHg of invasive blood pressure values respectively. In addition, 52% and 87% of diastolic arterial blood pressure (DAP) measurements were within 10 and 20 mmHg of invasive readings. High definition oscillometric readings did not meet ACVIM recommended limits for SAP.
The Doppler technique overestimated and the HDO device showed limited agreement with invasive blood pressure measurement in anaesthetized dogs. High definition oscillometry met most of the ACVIM requirements for MAP and DAP while the Doppler technique did not.
使用美国兽医内科学会(ACVIM)的验证标准,评估在麻醉犬中,高清示波法(HDO)和多普勒血压测量技术相对于有创血压测量的性能。
前瞻性临床研究。
20只客户拥有的犬。
使用插入足背动脉的导管和电子传感器测量有创血压。HDO测量的袖带放置部位为前臂中部或近端尾部,多普勒技术的袖带放置部位为胫骨远端。采用Bland-Altman法评估有创和无创血压测量之间的一致性。
分别仅有10%和34%的多普勒测量值在有创血压值的10 mmHg和20 mmHg范围内。多普勒设备未达到ACVIM血压测量设备的验证标准。HDO与有创血压测量技术之间,平均动脉压(MAP)的一致性最佳;分别有67%和95%的读数在有创血压值的10 mmHg和20 mmHg范围内。此外,分别有52%和87%的舒张压(DAP)测量值在有创读数的10 mmHg和20 mmHg范围内。高清示波法读数未达到ACVIM推荐的收缩压(SAP)限值。
在麻醉犬中,多普勒技术高估了血压,HDO设备与有创血压测量的一致性有限。高清示波法符合ACVIM对MAP和DAP的大部分要求,而多普勒技术则不符合。