Department of Radiation Oncology, Maastricht University Medical Center, Maastricht, The Netherlands.
Med Phys. 2012 Apr;39(4):1925-35. doi: 10.1118/1.3693049.
In gynecological radiotherapy with high dose rate (HDR)(192)Ir brachytherapy, the treatment complexity has increased due to improved optimization techniques and dose constraints. As a consequence, it has become more important to verify the dose delivery to the target and also to the organs at risk (e.g., the bladder). In vivo dosimetry, where dosimeters are placed in or on the patient, is one way of verifying the dose but until recently this was hampered by motion of the radiation detectors with respect to the source. The authors present a novel dosimetry method using a position sensitive radiation detector.
The prototype RADPOS system (Best Medical Canada) consists of a metal oxide field effect transistor (MOSFET) dosimeter coupled to a position-sensor, which deduces its 3D position in a magnetic field. To assess the feasibility of in vivo dosimetry based on the RADPOS system, different characteristics of the detector need to be investigated. Using a PMMA phantom, the positioning accuracy of the RADPOS system was quantified by comparing position readouts with the known position of the detector along the x and y-axes. RADPOS dose measurements were performed at various distances from a Nucletron(192)Ir source in a PMMA phantom to evaluate the energy dependence of the MOSFET. A sensitivity analysis was performed by calculating the dose after varying (1) the position of the RADPOS detector to simulate organ motion and (2) the position of the first dwell position to simulate errors in delivery. The authors also performed an uncertainty analysis to determine the action level (AL) that should be used during in vivo dosimetry.
Positioning accuracy is found to be within 1 mm in the 1-10 cm range from the origin along the x-axis (away from the transmitter), meeting the requirements for in vivo dosimetry. Similar results are obtained for the other axes. The ALs are chosen to take into account the total uncertainty on the measurements. As a consequence for in vivo dosimetry, it is determined that the RADPOS sensor, if placed, for example, in the bladder Foley balloon, would detect a 2 mm motion of the bladder, at a 5% chance of a false positive, with an AL limit of 9% of the dose delivered. The authors found that source position errors, caused by, e.g., a wrong first dwell position, are more difficult to detect; indeed, with our single RADPOS detector, positioned in the bladder, dwell position errors below 5 mm and resulting in a dose error within 10%, could be detected in the tandem but not in the colpostats. A possible solution to improve error detection is to use multiple MOSFETs to obtain multiple dose values.
In this study, the authors proposed a dosimetry procedure, based on the novel RADPOS system, to accurately determine the position of the radiation dosimeter with respect to the applicator. The authors found that it is possible to monitor the delivered dose in a point and compare it to the predetermined dose. This allows in principle the detection of problems such as bladder motion/filling or source mispositioning. Further clinical investigation is warranted.
在使用高剂量率(HDR)(192)Ir 近距离放射治疗的妇科放疗中,由于优化技术和剂量限制的改进,治疗复杂性增加了。因此,验证靶区和危及器官(如膀胱)的剂量输送变得更为重要。体内剂量学是一种验证剂量的方法,其中将剂量计放置在患者体内或体表,但直到最近,由于辐射探测器相对于源的运动,这种方法受到了阻碍。作者提出了一种使用位置敏感辐射探测器的新型剂量学方法。
原型 RADPOS 系统(Best Medical Canada)由金属氧化物场效应晶体管(MOSFET)剂量计与位置传感器耦合而成,该位置传感器可以推断出磁场中其三维位置。为了评估基于 RADPOS 系统的体内剂量学的可行性,需要研究探测器的不同特性。使用 PMMA 体模,通过比较探测器在 x 和 y 轴上的已知位置与位置读数,定量评估 RADPOS 系统的定位精度。在 PMMA 体模中,在距 Nucletron(192)Ir 源不同距离处进行 RADPOS 剂量测量,以评估 MOSFET 的能量依赖性。通过改变(1)RADPOS 探测器的位置以模拟器官运动,(2)第一驻留点的位置以模拟输送误差,进行了灵敏度分析。作者还进行了不确定度分析,以确定体内剂量学中应使用的行动水平(AL)。
发现,在距原点 1-10cm 范围内,沿 x 轴(远离发射器)的位置精度在 1mm 以内,满足体内剂量学的要求。在其他轴上也得到了类似的结果。AL 是根据测量的总不确定度选择的。因此,对于体内剂量学,如果将 RADPOS 传感器放置在例如膀胱 Foley 球囊中,那么在膀胱有 5%的假阳性概率时,将检测到 2mm 的膀胱运动,其 AL 限值为所输送剂量的 9%。作者发现,由于例如第一个驻留点的位置错误等原因导致的源位置误差更难检测;实际上,使用我们的单个 RADPOS 探测器,放置在膀胱中,如果驻留点误差小于 5mm 且导致剂量误差小于 10%,则可以在 tandem 中检测到,但不能在 colpostats 中检测到。提高误差检测能力的一种可能方法是使用多个 MOSFET 来获得多个剂量值。
在这项研究中,作者提出了一种基于新型 RADPOS 系统的剂量学方法,用于精确确定辐射剂量计相对于施源器的位置。作者发现,可以监测到一个点的输送剂量,并将其与预定剂量进行比较。这原则上允许检测到诸如膀胱运动/充盈或源定位错误等问题。需要进一步的临床研究。