The in vivo dosimetry tool, RADPOS, has been modified to include a metal oxide-silicon semiconductor field effect transistor (MOSFET) array with an electromagnetic positioning sensor. This allows dose monitoring at five points rather than just at single dose point as in the other versions of the device. The detector has been used in a clinical trial, which is the first to measure both urethral dose and internal motion concurrently during permanent seed implantation for prostate brachytherapy using a single probe.
The RADPOS detector was secured inside a Foley catheter inside the patient's urethra. Spatial coordinates of the RADPOS detector were read every 0.5s, and the timing of events such as needle insertion was noted. The MOSFET readings were taken over two 10-min periods; once all seeds had been implanted both before and after the transrectal ultrasound (TRUS), the probe was removed. Measurements were completed for 16 patients.
Maximum integral dose in the prostatic urethral ranged from 89 to 195Gy, and dose varied from -66% to 36% depending on the rectal probe position. The change in position of the RADPOS sensor owing to the removal of the TRUS probe ranged from 1.4 to 9.7mm.
The modified RADPOS detector with MOSFET array is able to provide real-time dose information, which can be used to monitor dose rates while implantation is performed and to estimate the total integrated dose. Changes in position including those owing to the TRUS probe can be significant and should be quantified to evaluate the influence on dose distributions.
体内剂量测定工具RADPOS已进行改进,纳入了带有电磁定位传感器的金属氧化物半导体场效应晶体管(MOSFET)阵列。这使得能够在五个点进行剂量监测,而不像该设备的其他版本那样仅在单个剂量点进行监测。该探测器已用于一项临床试验,这是首次在前列腺近距离治疗的永久性种子植入过程中,使用单个探头同时测量尿道剂量和内部运动。
将RADPOS探测器固定在患者尿道内的Foley导管内。每0.5秒读取一次RADPOS探测器的空间坐标,并记录诸如针插入等事件的时间。在两个10分钟时间段内获取MOSFET读数;在经直肠超声(TRUS)前后所有种子都植入完成后,将探头取出。对16名患者完成了测量。
前列腺尿道中的最大积分剂量范围为89至195Gy,剂量根据直肠探头位置在-66%至36%之间变化。由于取出TRUS探头,RADPOS传感器的位置变化范围为1.4至9.7mm。
带有MOSFET阵列的改进型RADPOS探测器能够提供实时剂量信息,可用于在植入过程中监测剂量率并估计总积分剂量。包括因TRUS探头引起的位置变化可能很大,应进行量化以评估对剂量分布的影响。