Suppr超能文献

使用金属氧化物半导体场效应晶体管探测器对高剂量率近距离放射治疗的计划剂量学进行验证。

Verification of the plan dosimetry for high dose rate brachytherapy using metal-oxide-semiconductor field effect transistor detectors.

作者信息

Qi Zhen-yu, Deng Xiao-wu, Huang Shao-min, Lu Jie, Lerch Michael, Cutajar Dean, Rosenfeld Anatoly

机构信息

State Key Laboratory of Oncology in Southern China, Guangzhou 510060, China.

出版信息

Med Phys. 2007 Jun;34(6):2007-13. doi: 10.1118/1.2736288.

Abstract

The feasibility of a recently designed metal-oxide-semiconductor field effect transistor (MOSFET) dosimetry system for dose verification of high dose rate (HDR) brachytherapy treatment planning was investigated. MOSFET detectors were calibrated with a 0.6 cm3 NE-2571 Farmer-type ionization chamber in water. Key characteristics of the MOSFET detectors, such as the energy dependence, that will affect phantom measurements with HDR 192Ir sources were measured. The MOS-FET detector was then applied to verify the dosimetric accuracy of HDR brachytherapy treatments in a custom-made water phantom. Three MOSFET detectors were calibrated independently, with the calibration factors ranging from 0.187 to 0.215 cGy/mV. A distance dependent energy response was observed, significant within 2 cm from the source. The new MOSFET detector has a good reproducibility (<3%), small angular effect (<2%), and good dose linearity (R2=1). It was observed that the MOSFET detectors had a linear response to dose until the threshold voltage reached approximately 24 V for 192Ir source measurements. Further comparison of phantom measurements using MOSFET detectors with dose calculations by a commercial treatment planning system for computed tomography-based brachytherapy treatment plans showed that the mean relative deviation was 2.2 +/- 0.2% for dose points 1 cm away from the source and 2.0 +/- 0.1% for dose points located 2 cm away. The percentage deviations between the measured doses and the planned doses were below 5% for all the measurements. The MOSFET detector, with its advantages of small physical size and ease of use, is a reliable tool for quality assurance of HDR brachytherapy. The phantom verification method described here is universal and can be applied to other HDR brachytherapy treatments.

摘要

研究了一种最近设计的金属氧化物半导体场效应晶体管(MOSFET)剂量测定系统用于高剂量率(HDR)近距离放射治疗计划剂量验证的可行性。MOSFET探测器在水中用0.6 cm³的NE - 2571 Farmer型电离室进行校准。测量了MOSFET探测器的关键特性,如能量依赖性,这些特性会影响使用HDR 192Ir源进行模体测量。然后将MOS - FET探测器应用于在定制水模体中验证HDR近距离放射治疗的剂量准确性。三个MOSFET探测器独立校准,校准因子范围为0.187至0.215 cGy/mV。观察到与距离相关的能量响应,在距源2 cm范围内显著。新型MOSFET探测器具有良好的重现性(<3%)、小角度效应(<2%)和良好的剂量线性(R² = 1)。观察到对于192Ir源测量,直到阈值电压达到约24 V之前,MOSFET探测器对剂量具有线性响应。使用MOSFET探测器进行的模体测量与基于计算机断层扫描的近距离放射治疗计划的商业治疗计划系统的剂量计算的进一步比较表明,距离源1 cm处的剂量点平均相对偏差为2.2±0.2%,距离源2 cm处的剂量点平均相对偏差为2.0±0.1%。所有测量中,测量剂量与计划剂量之间的百分比偏差均低于5%。MOSFET探测器具有物理尺寸小和使用方便的优点,是HDR近距离放射治疗质量保证的可靠工具。这里描述的模体验证方法具有通用性,可应用于其他HDR近距离放射治疗。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验