Department of Movement and Sports Sciences, Ghent University, B-9000 Ghent, Belgium.
J Sci Med Sport. 2012 Sep;15(5):418-24. doi: 10.1016/j.jsams.2012.02.002. Epub 2012 Apr 6.
To test the validity and intra-instrument reliability of the HJ-203 Omron pedometer when worn in different positions, under controlled and free-living conditions.
Cross-sectional.
Forty healthy adults (20 men, 29.5±7.7 years) participated in three controlled tests, i.e. 20 step test, stair climbing (up and down), and treadmill walking (five bouts at different speeds). All participants wore a HJ-203 pedometer in the pants pocket, in a carrier bag, and around the neck. Fifty-four adults (23 men, 33.9±11.1 years) participated in a free-living conditions test, wearing a HJ-203 in the pants pocket and around the neck during one day.
During controlled tests, absolute percentage error ranged between 0.1% and 14.0%. Accuracy was influenced by wearing position (p<0.001), walking speed (p<0.001), and wearing position×walking speed (p=0.001). Accuracy was poor for pedometers worn in the pants pocket, especially at slower speeds; and best when worn around the neck. During free-living conditions, APE ranged between 30.7% and 36.9% and did not differ statistically between pants pocket and neck position. Intra-instrument reliability varied for controlled tests (ICC=0.14-0.96) and was acceptable during free-living conditions (ICC=0.94).
The HJ-203 Omron pedometer showed acceptable accuracy for all wearing positions during stairs walking and treadmill walking at higher speeds; but limited accuracy during free-living conditions when worn at non-traditional wearing positions (necklace and pants pocket). Reliability was acceptable during treadmill walking at higher speeds and free-living conditions.
在受控和自由生活条件下,测试 HJ-203 欧姆龙计步器佩戴在不同位置时的有效性和仪器内信度。
横断面研究。
40 名健康成年人(20 名男性,29.5±7.7 岁)参与了三项对照测试,即 20 步测试、爬楼梯(上下)和跑步机步行(以不同速度进行五次)。所有参与者均在裤子口袋、携带袋和脖子上佩戴 HJ-203 计步器。54 名成年人(23 名男性,33.9±11.1 岁)在一天内佩戴 HJ-203 计步器,在裤子口袋和脖子上各佩戴一个计步器,参与自由生活条件测试。
在对照测试中,绝对百分比误差在 0.1%至 14.0%之间。准确性受佩戴位置(p<0.001)、行走速度(p<0.001)和佩戴位置×行走速度(p=0.001)的影响。在裤子口袋中佩戴计步器的准确性较差,尤其是在较慢的速度下;而在脖子上佩戴的准确性最佳。在自由生活条件下,APE 在 30.7%至 36.9%之间,裤子口袋和脖子位置之间无统计学差异。仪器内信度在对照测试中变化(ICC=0.14-0.96),在自由生活条件下可接受(ICC=0.94)。
HJ-203 欧姆龙计步器在楼梯行走和较高速度跑步机行走时,所有佩戴位置的准确性均可接受;但在非传统佩戴位置(项链和裤子口袋)时,自由生活条件下的准确性有限。在较高速度跑步机行走和自由生活条件下,可靠性可接受。