This article describes a case of late locking mechanism failure and disassociation of a Pinnacle acetabular cup (DePuy, Warsaw, Indiana) and Marathon polyethylene liner (DePuy) 53 months after routine primary total hip arthroplasty in an active patient. Following an uncomplicated initial postoperative recovery, the patient felt a pop while kneeling for gardening activities. The patient had no prodromal symptoms but reported pain with range of motion and weight bearing following the episode. Radiographs appeared to show a fractured liner with intact acetabular and femoral components. Intraoperatively, the liner was disassociated and dislocated inferior to the acetabulum, and 3 consecutive antirotational tines were sheared off the liner at their bases. The acetabular cup and femoral stem were well fixed. Extensive metallosis existed from the femoral head and acetabular shell articulating in the absence of a liner. The patient underwent revision of the acetabular cup, polyethylene liner, and femoral head due to the concern for osteolysis given the extensive metallosis. Revision of the acetabular cup and liner were performed with no further complications. No conclusions could be made as to the exact mechanism of failure following laboratory analysis of the retrieved polyethylene liner. Several possible mechanisms of failure are possible. Although cases of acute disassociation of this system have been reported, this is the first article to our knowledge to report failure at such a late postoperative time.
本文描述了一例在一名活跃患者进行常规初次全髋关节置换术后53个月时,Pinnacle髋臼杯(迪普伊公司,印第安纳州华沙)和马拉松聚乙烯内衬(迪普伊公司)出现晚期锁定机制故障和分离的病例。术后初期恢复顺利,患者在跪着进行园艺活动时感到一声“啪”响。患者无前驱症状,但称该事件后活动及负重时疼痛。X线片显示内衬骨折,髋臼和股骨组件完好。术中,内衬分离并脱位于髋臼下方,连续3个抗旋转齿在其基部从内衬上被剪断。髋臼杯和股骨干固定良好。由于没有内衬,股骨头与髋臼杯之间的关节出现广泛的金属沉着病。鉴于广泛的金属沉着病,考虑到骨溶解问题,患者接受了髋臼杯、聚乙烯内衬和股骨头的翻修。髋臼杯和内衬的翻修未出现进一步并发症。对取出的聚乙烯内衬进行实验室分析后,无法确定确切的失效机制。可能存在几种失效机制。虽然已有该系统急性分离的病例报道,但据我们所知,本文是第一篇报道术后如此长时间出现故障的文章。