Suppr超能文献

MRI 平扫轴向容积屏气检查在流行病学研究中测量胸腹部主动脉直径:一项验证研究。

Thoracic and abdominal aortic diameter measurement by MRI using plain axial volumetric interpolated breath-hold examination in epidemiologic research: a validation study.

机构信息

Institute of Diagnostic Radiology and Neuroradiology, University Medicine Greifswald, Germany.

出版信息

Acad Radiol. 2012 Aug;19(8):1011-7. doi: 10.1016/j.acra.2012.03.012. Epub 2012 Apr 13.

Abstract

RATIONALE AND OBJECTIVES

This study evaluates the validity and reliability of measuring the diameters of the thoracic and abdominal aorta from plain volumetric interpolated breath-hold examination (VIBE) images.

MATERIALS AND METHODS

The study included 50 male subjects from the population-based Study of Health in Pomerania. They underwent imaging of the thoracic and abdominal aorta at 1.5 Tesla using a contrast-enhanced magnetic resonance angiography (CE-MRA) and plain VIBE sequence. Diameters were measured at five predefined anatomic sites from reformatted orthogonal CE-MRA images and axial plain VIBE images. The measurements were validated using Pearson correlation and Bland-Altman analysis. The Bland-Altman method was also used to assess reliability.

RESULTS

Comparison of the diameters measured from CE-MRA and VIBE images revealed strong correlation for the ascending, descending, suprarenal, and infrarenal aorta with r = 0.95 (P < .0001), r = 0.88 (P < .0001), 0.92 (P < .0001), and 0.87 (P < .0001), respectively. Measurement for the aortic arch was moderately correlated with r = 0.78 (P < .0001). Mean bias did not exceed 0.1 cm (6%). The 95% limits of agreement (LOA) were less than 0.5 cm (15%). Intra- and interobserver agreement showed a mean bias of less than 2%; the 95% LOA were less than 11%.

CONCLUSIONS

Axial measurement of the diameters of the thoracic and abdominal aorta using a plain axial VIBE sequence is highly valid and reliable, making it suitable for use in epidemiologic research.

摘要

目的

本研究评估了从平扫容积内插屏气检查(VIBE)图像测量胸、腹主动脉直径的有效性和可靠性。

材料与方法

本研究纳入了来自基于人群的波罗的海健康研究中的 50 名男性受试者。他们在 1.5T 磁共振仪上进行了胸、腹主动脉对比增强磁共振血管造影(CE-MRA)和 VIBE 平扫序列的成像。在重建成角 CE-MRA 图像和轴向 VIBE 平扫图像上,在五个预设的解剖部位测量直径。使用 Pearson 相关和 Bland-Altman 分析对测量值进行验证。Bland-Altman 法也用于评估可靠性。

结果

CE-MRA 和 VIBE 图像测量的直径之间的比较显示,升主动脉、降主动脉、肾上极和肾下极的相关性很强,r 值分别为 0.95(P <.0001)、r = 0.88(P <.0001)、0.92(P <.0001)和 0.87(P <.0001)。主动脉弓的测量结果为中度相关,r = 0.78(P <.0001)。平均偏差不超过 0.1cm(6%)。95%的一致性区间(LOA)小于 0.5cm(15%)。内观察者和外观察者的一致性显示平均偏差小于 2%;95%的 LOA 小于 11%。

结论

使用平扫轴向 VIBE 序列测量胸、腹主动脉的直径具有高度的有效性和可靠性,适用于流行病学研究。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验