Institute of Diagnostic Radiology and Neuroradiology, University Medicine Greifswald, Ferdinand-Sauerbruch-Straße 1, 17475 Greifswald, Germany.
Acad Radiol. 2013 Aug;20(8):1004-8. doi: 10.1016/j.acra.2013.03.014.
To assess the validity and reliability of measuring mean aortic wall thickness (MAWT) of the ascending and descending aorta using cine steady-state free precession (SSFP) imaging compared to dark blood (DB) imaging.
DB and SSFP images of the thoracic aorta acquired at 1.5 T in 50 volunteers (26 women, 24 men; mean age: 50.2 ± 13.1 years) were used. MAWT was calculated on DB and SSFP images for the ascending and descending aorta at the level of the right pulmonary artery by two independent observers. Validity was assessed using Bland-Altman analysis, Passing-Bablok regression, and Spearman correlation. Reliability was assessed using Bland-Altman analysis and intraclass coefficients (ICCs).
The mean MAWT of the ascending aorta on DB and SSFP images was 1.89 ± 0.21 mm and 1.87 ± 0.20 mm. The measurements for the descending aorta were 1.60 ± 0.22 and 1.63 ± 0.20 mm, respectively. Comparison of DB and SSFP measurements revealed a mean bias of 1.3% (95% limits of agreement (LOA): -7.9, 10.5%) for the ascending and of -2.1% (LOA: -10.5, 6.3%) for the descending aorta. The corresponding regression equation was y = 0.042 + 0.960 × (r = 0.91; P < .0001) and y = 0.118 + 0.939 × (r = 0.95; P < .0001), respectively. Intra- and interobserver variability showed a mean bias of less than 2.0% and LOA of less than ±15.0%. ICCs were greater than or equal to 0.85.
MAWT determination in the ascending and descending aorta using cine SSFP sequences is highly valid and reliable compared to DB imaging.
评估使用电影稳态自由进动(cine SSFP)成像测量升主动脉和降主动脉平均主动脉壁厚度(MAWT)与黑血(DB)成像相比的有效性和可靠性。
在 1.5 T 扫描仪上对 50 名志愿者(26 名女性,24 名男性;平均年龄:50.2 ± 13.1 岁)进行胸部主动脉的 DB 和 cine SSFP 图像采集。由两名独立观察者在右肺动脉水平对升主动脉和降主动脉的 DB 和 cine SSFP 图像进行 MAWT 计算。使用 Bland-Altman 分析、Passing-Bablok 回归和 Spearman 相关分析评估有效性。使用 Bland-Altman 分析和组内相关系数(ICC)评估可靠性。
DB 和 cine SSFP 图像上的升主动脉 MAWT 平均值分别为 1.89 ± 0.21 mm 和 1.87 ± 0.20 mm。降主动脉的测量值分别为 1.60 ± 0.22 和 1.63 ± 0.20 mm。DB 和 cine SSFP 测量值的比较显示,升主动脉的平均偏差为 1.3%(95%一致性区间(LOA):-7.9,10.5%),降主动脉的平均偏差为-2.1%(LOA:-10.5,6.3%)。相应的回归方程分别为 y = 0.042 + 0.960 ×(r = 0.91;P <.0001)和 y = 0.118 + 0.939 ×(r = 0.95;P <.0001)。观察者内和观察者间的变异性显示平均偏差小于 2.0%,LOA 小于±15.0%。ICC 大于或等于 0.85。
与 DB 成像相比,使用电影 SSFP 序列测量升主动脉和降主动脉的 MAWT 具有很高的有效性和可靠性。