Department of Linguistics, Leopold-Franzens University, Innsbruck, Austria.
Otol Neurotol. 2012 Jun;33(4):618-22. doi: 10.1097/MAO.0b013e3182536909.
A prospective, single-subjects repeated-measures study design was used to compare speech understanding performance with the 2 APs acutely and after 6 weeks' acclimatization time.
Tertiary referral center.
Thirteen experienced unilateral German-speaking active middle ear implant users with either mixed or conductive hearing loss were enrolled.
Directional audio processor.
SNR50 obtained from 4 different noise conditions.
SNR50s with the directional AP were lower (better) than with the omnidirectional AP in all noise conditions, although there were no changes in aided PTA thresholds. In acute testing, the mean SNR50 showed a significant directional advantage (DA) for the directional AP in the S0N0 condition of 1.3 dB, in S0NVSB 2.3 dB, in S0Ncont 3.1 dB, and in S0N180 4.5 dB. As expected, the largest DA was found in the S0N180 condition. No DA was expected in the S0N0 condition but was present. After 6 weeks of acclimatization time, no significant change from the acute testing was found, suggesting that patients experienced improved performance with the directional AP, even without acclimatization time. There was no significant change in subjective questionnaire outcomes.
The directional AP yields immediately improved speech understanding in noise.
1)通过在 4 种不同噪声条件下比较使用全向和定向麦克风音频处理器(AP)时 50%正确单词理解的信噪比(SNR50),评估噪声下的言语理解能力。2)比较两种处理器的主观言语理解能力、空间听觉能力和定性听力表现。
前瞻性、单受试者重复测量研究设计用于在急性和 6 周适应期后比较两种 AP 的言语理解性能。
三级转诊中心。
13 名经验丰富的单侧德语活跃中耳植入患者,伴有混合性或传导性听力损失。
定向音频处理器。
从 4 种不同噪声条件获得的 SNR50。
在所有噪声条件下,定向 AP 的 SNR50 均低于(更好)全向 AP,尽管辅助 PTA 阈值没有变化。在急性测试中,定向 AP 在 S0N0 条件下的平均 SNR50 表现出显著的定向优势(DA)为 1.3dB,在 S0NVSB 为 2.3dB,在 S0Ncont 为 3.1dB,在 S0N180 为 4.5dB。正如预期的那样,S0N180 条件下的 DA 最大。S0N0 条件下不应出现 DA,但实际上存在。经过 6 周的适应期,急性测试没有发现显著变化,表明患者即使没有适应期,也能从定向 AP 中获得更好的表现。主观问卷调查结果没有显著变化。
定向 AP 可立即提高噪声下的言语理解能力。