The new Certas™ shunt for the treatment of hydrocephalus has seven standard pressure settings that according to the manufacturer range from 36 to 238 mmH2O, and an additional "Virtual Off" setting with an opening pressure >400 mmH2O. Information on actual pressure response and reliability of shunt performance is important in clinical application, especially the "Virtual Off" setting as a non-surgical replacement for shunt ligation. The objective of this study was to evaluate the in-vitro hydrodynamic performance of the Certas™ shunt.
Six new Certas™ shunts with proximal and distal catheters were tested with an automated, computerized test system that raised the pressure from zero to a maximum pressure and back to zero at each valve setting. Opening pressure and flow resistance were determined.
For settings 1-7 the measured opening pressure range was 26 to 247 mmH2O, and the mean change in opening pressure for a one-step adjustment was between 33 and 38 mmH2O. For setting 8 ("Virtual Off") the measured mean opening pressure was 494 ± 34 mmH2O (range 451 to 556 mmH2O). The mean outflow resistance was 7.0 mmHg/ml/min (outflow conductance 17.9 μl/s/kPa).
The six shunts had similar characteristics and closely matched the manufacturer's specifications for opening pressure at settings 1-7. The opening pressure for the "Virtual Off" setting was nearly 500 mmH2O, which is 100 mmH2O higher than the manufacturer's specification of ">400" and should be functionally off for most patients with communicating hydrocephalus. Clinical studies are needed to evaluate if the CSF dynamic profile persists after implantation in patients.
新型 Certas™ 分流器用于治疗脑积水,有七个标准压力设置,根据制造商的说法,范围从 36 到 238mmH2O,还有一个额外的“虚拟关闭”设置,其开启压力>400mmH2O。了解实际压力响应和分流器性能的可靠性对于临床应用非常重要,特别是“虚拟关闭”设置作为分流器结扎的非手术替代方法。本研究旨在评估 Certas™ 分流器的体外流体动力学性能。
使用自动化计算机测试系统对 6 个新的 Certas™ 分流器及其近端和远端导管进行测试,该系统在每个阀门设置下将压力从零升高到最大压力,然后再回到零。测定开启压力和流动阻力。
在设置 1-7 时,测量的开启压力范围为 26 至 247mmH2O,一步调整时开启压力的平均变化在 33 至 38mmH2O 之间。对于设置 8(“虚拟关闭”),测量的平均开启压力为 494±34mmH2O(范围为 451 至 556mmH2O)。平均流出阻力为 7.0mmHg/ml/min(流出导纳为 17.9μl/s/kPa)。
六个分流器具有相似的特征,与制造商规定的设置 1-7 时的开启压力非常匹配。“虚拟关闭”设置的开启压力接近 500mmH2O,比制造商规定的“>400”高 100mmH2O,对于大多数交通性脑积水患者来说应该是功能关闭的。需要进行临床研究以评估在患者体内植入后 CSF 动态特征是否持续。