Assessment of the intraobserver and interobserver reliability of a communicating vessels volumeter to measure wrist-hand volume.

机构信息

Department of Vascular Surgery, Paulista School of Medicine, Federal University of São Paulo (UNIFESP), São Paulo, Brazil.

出版信息

Phys Ther. 2012 Oct;92(10):1329-37. doi: 10.2522/ptj.20110169. Epub 2012 Jun 7.

Abstract

BACKGROUND

Traditional volumetry based on Archimedes' principle is the gold standard for the measurement of limb volume, but the routine use of this technique is discouraged because of several disadvantages.

OBJECTIVE

The purpose of this study was to evaluate intraobserver and interobserver reliability of direct measurements of wrist-hand volume using a new communicating vessels volumeter based on Pascal's law.

DESIGN

A reliability study was conducted.

METHODS

To evaluate the reliability of the communicating vessels volumeter in generating measurements, 30 hands of 15 participants (9 women, 6 men) were measured 3 times each by 3 observers, totaling 270 volumetric results.

RESULTS

Measurement time was short (X =3 minutes 42 seconds). The intraclass correlation coefficient (ICC) was .9977 for observer 1 and .9976 for observers 2 and 3. The interobserver ICC was .9998. The standard error of measurement was about 3 mL for all observers; the interobserver result was 1 mL. The interrater coefficient of variance (CV) was 1.15% for the series of 9 measurements collected for each segment; the intrarater CV was 1.20%. Limitations No swollen hands were measured, and measurements were not compared with the gold standard technique. Thus, accuracy of the new volumeter was not determined in this study.

CONCLUSION

A new device has been developed for plethysmography of the extremities, and the results of its use to measure the volume of the wrist-hand segment were reliable in both intraobserver and interobserver analyses.

摘要

背景

基于阿基米德原理的传统体积测量法是测量肢体体积的金标准,但由于存在多种缺点,不鼓励常规使用该技术。

目的

本研究旨在评估基于帕斯卡定律的新型连通容器体积计对手腕-手部体积进行直接测量的观察者内和观察者间可靠性。

设计

可靠性研究。

方法

为了评估连通容器体积计在产生测量结果方面的可靠性,由 3 名观察者对 15 名参与者(9 名女性,6 名男性)的 30 只手进行了 3 次测量,总共有 270 个体积结果。

结果

测量时间短(X = 3 分 42 秒)。观察者 1 的组内相关系数(ICC)为.9977,观察者 2 和 3 的 ICC 为.9976。观察者间的 ICC 为.9998。所有观察者的测量标准误差约为 3 mL;观察者间的结果为 1 mL。系列 9 次测量中每个节段的观察者间变异系数(CV)为 1.15%;观察者内的 CV 为 1.20%。局限性:未测量肿胀的手,且测量结果未与金标准技术进行比较。因此,本研究未确定新体积计的准确性。

结论

已开发出一种用于四肢体积描记术的新设备,并且其用于测量腕手部体积的使用结果在观察者内和观察者间分析中均具有可靠性。