Precision and accuracy measurement of radiostereometric analysis applied to movement of the sacroiliac joint.

机构信息

Division of Surgery and Clinical Neuroscience, Department of Orthopaedics, Oslo University Hospital, Pb 4950 Nydalen, 0424 Oslo, Norway.

出版信息

Clin Orthop Relat Res. 2012 Nov;470(11):3187-94. doi: 10.1007/s11999-012-2413-5. Epub 2012 Jun 14.

Abstract

BACKGROUND

Different techniques have been used to quantify the movement of sacroiliac (SI) joints. These include radiostereometric analysis (RSA), but the accuracy and precision of this method have not been properly evaluated and it is unclear how many markers are required and where they should be placed to achieve proper accuracy and precision.

PURPOSE

The purpose of this study was to test accuracy and precision of RSA, applied to the SI joint, in a phantom model and in patients.

METHODS

We used a plastic phantom attached to a micrometer to obtain a true value of the movement of the SI joint and compared this value with the measured value obtained by RSA; the difference represented the accuracy. The precision of the system was measured by double examination in the phantom and in six patients, and was expressed by a limit of significance (LOS). We analyzed different marker distributions to find optimal marker placement and number of markers needed.

RESULTS

The accuracy was high and we identified no systematic errors. The precision of the phantom was high with a LOS less than 0.25° and 0.16 mm for all directions, and in patients, the precision was less than 0.71° for rotations and 0.47 mm translations. No markers were needed in the pubic symphysis to obtain good precision.

CONCLUSIONS

The accuracy and precision are high when RSA is used to measure movement in the SI joint and support the use of RSA in research of SI joint motion.

摘要

背景

已经使用了不同的技术来量化骶髂(SI)关节的运动。这些技术包括放射立体测量分析(RSA),但该方法的准确性和精密度尚未得到适当评估,并且不清楚需要多少个标记以及应将其放置在何处以达到适当的准确性和精密度。

目的

本研究的目的是在体模和患者中测试 RSA 对 SI 关节运动的准确性和精密度。

方法

我们使用附接到千分尺的塑料体模来获得 SI 关节运动的真实值,并将该值与通过 RSA 获得的测量值进行比较;差异代表准确性。通过在体模和六名患者中进行双重检查来测量系统的精密度,并通过显著性限(LOS)表示。我们分析了不同的标记分布,以找到最佳的标记放置和所需的标记数量。

结果

准确性很高,我们没有发现系统误差。体模的精密度很高,所有方向的 LOS 均小于 0.25°和 0.16mm,而在患者中,旋转的精密度小于 0.71°,平移的精密度小于 0.47mm。在耻骨联合处不需要标记即可获得良好的精密度。

结论

当使用 RSA 测量 SI 关节运动时,准确性和精密度都很高,这支持了 RSA 在 SI 关节运动研究中的应用。

相似文献

引用本文的文献

9
Experimental evaluation of precision and accuracy of RSA in the lumbar spine.RSA 在腰椎中的精度和准确性的实验评估。
Eur Spine J. 2021 Jul;30(7):2060-2068. doi: 10.1007/s00586-020-06672-5. Epub 2020 Dec 4.

本文引用的文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验