Kelting T, Searles B, Darling E
Department of Cardiovascular Perfusion, College of Health Professions, SUNY Upstate Medical University, Syracuse, NY 13210, USA.
Perfusion. 2012 Jul;27(4):345-51. doi: 10.1177/0267659112446526.
The ideal location of air bubble detector (ABD) placement on the cardiopulmonary bypass (CPB) circuit is debatable. There is, however, very little data characterizing the prevalence of specific ABD placement preferences by perfusionists. Therefore, the purpose of this study was to survey the perfusion community to collect data describing the primary locations of air bubble detector placement on the CPB circuit. In June 2011, an 18-question on-line survey was conducted. Completed surveys were received from 627 participants. Of these, analysis of the responses from the 559 certified clinical perfusionists (CCP) was performed. The routine use of ABD during CPB was reported by 96.8% of CCPs. Of this group, specific placement of the bubble detector is as follows: distal to the venous reservoir outlet (35.6%), between the arterial pump and oxygenator (3.8%), between the oxygenator and arterial line filter (35.1%), distal to the arterial line filter (ALF) (23.6%), and other (1.8%). Those placing the ABD distal to the venous reservoir predominately argued that an emptied venous reservoir was the most likely place to introduce air into the circuit. Those who placed the ABD between the oxygenator and the arterial line filter commonly reasoned that this placement protects against air exiting the membrane. Those placing the ABD distal to the ALF (23.6%) cited that this location protects from all possible entry points of air. A recent false alarm event from an ABD during a case was reported by 36.1% of CCPs. This study demonstrates that the majority of CCPs use an ABD during the conduct of CPB. The placement of the ABD on the circuit, however, is highly variable across the perfusion community. A strong rationale for the various ABD placements suggests that the adoption of multiple ABD may offer the greatest comprehensive protection against air emboli.
体外循环(CPB)回路中气泡探测器(ABD)的理想放置位置存在争议。然而,关于灌注师对特定ABD放置偏好的普遍性的数据非常少。因此,本研究的目的是对灌注领域进行调查,以收集描述CPB回路上气泡探测器主要放置位置的数据。2011年6月,进行了一项包含18个问题的在线调查。共收到627名参与者完成的调查问卷。其中,对559名注册临床灌注师(CCP)的回复进行了分析。96.8%的CCP报告在CPB期间常规使用ABD。在这一组中,气泡探测器的具体放置位置如下:静脉储血器出口远端(35.6%)、动脉泵和氧合器之间(3.8%)、氧合器和动脉管路过滤器之间(35.1%)、动脉管路过滤器(ALF)远端(23.6%)以及其他位置(1.8%)。将ABD放置在静脉储血器远端的人主要认为排空的静脉储血器是最有可能将空气引入回路的地方。将ABD放置在氧合器和动脉管路过滤器之间的人通常认为这种放置方式可防止空气从膜中逸出。将ABD放置在ALF远端的人(23.6%)称该位置可防止空气从所有可能的进入点进入。36.1%的CCP报告在一次病例中近期发生了ABD误报警事件。本研究表明,大多数CCP在CPB过程中使用ABD。然而,在整个灌注领域中,ABD在回路上的放置差异很大。对各种ABD放置方式的有力解释表明,采用多个ABD可能提供针对空气栓塞的最大综合保护。