Newland Richard F, Baker Robert A, Mazzone Annette L, Valiyapurayil Vijaykumar N
Cardiac Surgery Research and Perfusion, Flinders Medical Centre, Bedford Park, South Australia ; Flinders University, Bedford Park, South Australia.
Cardiac Surgery Research and Perfusion, Flinders Medical Centre, Bedford Park, South Australia.
J Extra Corpor Technol. 2015 Sep;47(3):174-9.
Air bubble detectors (ABDs) are utilized during cardiopulmonary bypass (CPB) to protect against massive air embolism. Stockert (Munich, Germany) ABD quantify microbubbles >300 μm; however, their reliability has not been reported. The aim of this study was to assess the reliability of the microbubble data from the ABD with the SIII and S5 heart-lung machines. Microbubble counts from the ABD with the SIII (SIII ABD) and S5 (S5 ABD) were measured simultaneously with the emboli detection and classification (EDAC) quantifier in 12 CPB procedures using two EDAC detectors and two ABDs in series in the arterial line. Reliability was assessed by the Spearman correlation co-efficient (r) between measurements for each detector type, and between each ABD and EDAC detector for counts >300 μm. No correlation was found between the SIII ABD (r = .008, p = .793). A weak negative correlation was found with the S5 ABD (r = -.16, p < .001). A strong correlation was found between the EDAC detectors (SIII; r = .958, p < .001), (S5; r = .908, p < .001). With counts >300 μm, the SIII ABDs showed a correlation of small-medium effect size between EDAC detectors and ABD1 (r = .286, p < .001 [EDAC1], r = .347, p < .001 [EDAC2]). There was no correlation found between ABD2 and either EDAC detector (r = .003, p = .925 (EDAC1), r = .003, p = .929 [EDAC2]). A correlation between EDAC and the S5 ABD, was not able to be determined due to the low bubble count detected by the EDAC >300 μm. Both SIII ABD and S5 ABD were found to be unreliable for quantification of microbubble activity during CPB in comparison with the EDAC. These results highlight the importance of ensuring that data included in the CPB report is accurate and clinically relevant, and suggests that microbubble counts from devices such as the SIII ABD and S5 ABD should not be reported.
气泡探测器(ABD)在体外循环(CPB)期间用于防止大量空气栓塞。斯托克特(德国慕尼黑)的ABD可对直径大于300μm的微泡进行量化;然而,其可靠性尚未见报道。本研究的目的是评估使用SIII和S5心肺机时来自ABD的微泡数据的可靠性。在12例CPB手术中,使用两个串联在动脉管路中的栓子检测与分类(EDAC)探测器和两个ABD,同时测量SIII(SIII ABD)和S5(S5 ABD)的ABD的微泡计数。通过每种探测器类型测量值之间以及每个ABD与EDAC探测器对于直径大于300μm的计数之间的斯皮尔曼相关系数(r)来评估可靠性。未发现SIII ABD之间存在相关性(r = 0.008,p = 0.793)。发现与S5 ABD存在弱负相关(r = -0.16,p < 0.001)。在EDAC探测器之间发现强相关性(SIII;r = 0.958,p < 0.001),(S5;r = 0.908,p < 0.001)。对于直径大于300μm的计数,SIII ABD在EDAC探测器与ABD1之间显示出中小效应量的相关性(r = 0.286,p < 0.001 [EDAC1],r = 0.347,p < 0.001 [EDAC2])。未发现ABD2与任何一个EDAC探测器之间存在相关性(r = 0.003,p = 0.925 [EDAC1],r = 0.003,p = 0.929 [EDAC2])。由于EDAC检测到的直径大于300μm的气泡计数较低,无法确定EDAC与S5 ABD之间的相关性。与EDAC相比,发现SIII ABD和S5 ABD在CPB期间对微泡活性的量化均不可靠。这些结果突出了确保CPB报告中包含的数据准确且与临床相关的重要性,并表明不应报告来自SIII ABD和S5 ABD等设备的微泡计数。