Impact of iterative reconstruction on CNR and SNR in dynamic myocardial perfusion imaging in an animal model.
机构信息
Institut für Radiologie, Klinikum rechts der Isar der Technischen Universität München, Ismaningerstrasse 22, 81675, München, Deutschland.
出版信息
Eur Radiol. 2012 Dec;22(12):2654-61. doi: 10.1007/s00330-012-2525-z. Epub 2012 Jul 3.
OBJECTIVES
To evaluate a new iterative reconstruction (IR) algorithm for radiation dose, image quality (IQ), signal-to-noise-ratio (SNR), and contrast-to-noise-ratio (CNR) in multidetector computed tomography (MDCT) dynamic myocardial perfusion imaging (MPI).
METHODS
ECG-gated 256-slice MDCT dynamic MPI was performed in six pigs after subtotal balloon occlusion of one artery. Two 100 kVp protocols were compared: high dose (HD): 150 mAs; low dose (LD): 100 mAs. HD images were reconstructed with filtered back projection (FBP), LD images with FBP and different strengths of IR (L1, L4, and L7). IQ (5-point scale), SNR, and CNR (ischemic vs. normal myocardium) values derived from the HD (FBP) images and the different LD images were compared.
RESULTS
Mean SNR values for myocardium were 16.3, 11.3, 13.1, 17.1, and 28.9 for the HD, LD (FBP), LD (L1), LD (L4), and LD (L7) reconstructions, respectively. Mean CNR values were 8.9, 6.3, 7.8, 9.3, and 12.8. IQ was scored as 4.6, 3.3, 4.4, 4.7, and 3.4, respectively. A significant loss of IQ was observed for the LD (L7) images compared to the HD (FBP) images (P < 0.05).
CONCLUSION
Appropriate levels of iterative reconstruction can improve SNR and CNR, facilitating radiation dose savings in CT-MPI without influencing diagnostic quality.
KEY POINTS
Iterative reconstruction (IR) can reduce radiation dose in myocardial perfusion CT. Our study also demonstrated improvements in image quality (noise, SNR, and CNR). Dynamic CT-MPI could help determine the hemodynamic significance of coronary artery disease. With dynamic CT MPI, myocardial blood flow can be determined quantitatively.
目的
评估一种新的迭代重建(IR)算法在多排螺旋 CT 动态心肌灌注成像(MPI)中的辐射剂量、图像质量(IQ)、信噪比(SNR)和对比噪声比(CNR)。
方法
对 6 只猪进行了部分球囊闭塞一支动脉后的 ECG 门控 256 层 MDCT 动态 MPI。比较了两种 100 kVp 协议:高剂量(HD):150 mAs;低剂量(LD):100 mAs。HD 图像采用滤波反投影(FBP)重建,LD 图像采用 FBP 和不同强度的 IR(L1、L4 和 L7)重建。比较了 HD(FBP)图像和不同 LD 图像得出的 IQ(5 分制)、SNR 和 CNR(缺血与正常心肌)值。
结果
心肌的平均 SNR 值分别为 16.3、11.3、13.1、17.1 和 28.9,对应的是 HD、LD(FBP)、LD(L1)、LD(L4)和 LD(L7)重建。平均 CNR 值分别为 8.9、6.3、7.8、9.3 和 12.8。IQ 评分分别为 4.6、3.3、4.4、4.7 和 3.4。与 HD(FBP)图像相比,LD(L7)图像的 IQ 明显下降(P < 0.05)。
结论
适当水平的迭代重建可以提高 SNR 和 CNR,在不影响诊断质量的情况下,有助于 CT-MPI 中的辐射剂量节省。
关键点
迭代重建(IR)可降低心肌灌注 CT 的辐射剂量。我们的研究还表明,图像质量(噪声、SNR 和 CNR)得到了改善。动态 CT-MPI 有助于确定冠状动脉疾病的血流动力学意义。通过动态 CT MPI,可以定量确定心肌血流量。