Suppr超能文献

噪声空间非均匀性及统计图像重建对 CT 心肌灌注成像的影响。

Noise spatial nonuniformity and the impact of statistical image reconstruction in CT myocardial perfusion imaging.

机构信息

Department of Medical Physics, University of Wisconsin-Madison, Madison, WI, USA.

出版信息

Med Phys. 2012 Jul;39(7):4079-92. doi: 10.1118/1.4722983.

Abstract

PURPOSE

To achieve high temporal resolution in CT myocardial perfusion imaging (MPI), images are often reconstructed using filtered backprojection (FBP) algorithms from data acquired within a short-scan angular range. However, the variation in the central angle from one time frame to the next in gated short scans has been shown to create detrimental partial scan artifacts when performing quantitative MPI measurements. This study has two main purposes. (1) To demonstrate the existence of a distinct detrimental effect in short-scan FBP, i.e., the introduction of a nonuniform spatial image noise distribution; this nonuniformity can lead to unexpectedly high image noise and streaking artifacts, which may affect CT MPI quantification. (2) To demonstrate that statistical image reconstruction (SIR) algorithms can be a potential solution to address the nonuniform spatial noise distribution problem and can also lead to radiation dose reduction in the context of CT MPI.

METHODS

Projection datasets from a numerically simulated perfusion phantom and an in vivo animal myocardial perfusion CT scan were used in this study. In the numerical phantom, multiple realizations of Poisson noise were added to projection data at each time frame to investigate the spatial distribution of noise. Images from all datasets were reconstructed using both FBP and SIR reconstruction algorithms. To quantify the spatial distribution of noise, the mean and standard deviation were measured in several regions of interest (ROIs) and analyzed across time frames. In the in vivo study, two low-dose scans at tube currents of 25 and 50 mA were reconstructed using FBP and SIR. Quantitative perfusion metrics, namely, the normalized upslope (NUS), myocardial blood volume (MBV), and first moment transit time (FMT), were measured for two ROIs and compared to reference values obtained from a high-dose scan performed at 500 mA.

RESULTS

Images reconstructed using FBP showed a highly nonuniform spatial distribution of noise. This spatial nonuniformity led to large fluctuations in the temporal direction. In the numerical phantom study, the level of noise was shown to vary by as much as 87% within a given image, and as much as 110% between different time frames for a ROI far from isocenter. The spatially nonuniform noise pattern was shown to correlate with the source trajectory and the object structure. In contrast, images reconstructed using SIR showed a highly uniform spatial distribution of noise, leading to smaller unexpected noise fluctuations in the temporal direction when a short scan angular range was used. In the numerical phantom study, the noise varied by less than 37% within a given image, and by less than 20% between different time frames. Also, the noise standard deviation in SIR images was on average half of that of FBP images. In the in vivo studies, the deviation observed between quantitative perfusion metrics measured from low-dose scans and high-dose scans was mitigated when SIR was used instead of FBP to reconstruct images.

CONCLUSIONS

(1) Images reconstructed using FBP suffered from nonuniform spatial noise levels. This nonuniformity is another manifestation of the detrimental effects caused by short-scan reconstruction in CT MPI. (2) Images reconstructed using SIR had a much lower and more uniform noise level and thus can be used as a potential solution to address the FBP nonuniformity. (3) Given the improvement in the accuracy of the perfusion metrics when using SIR, it may be desirable to use a statistical reconstruction framework to perform low-dose dynamic CT MPI.

摘要

目的

为了在 CT 心肌灌注成像(MPI)中实现高时间分辨率,通常使用从短扫描角度范围内采集的数据通过滤波反投影(FBP)算法来重建图像。然而,门控短扫描中从一帧到下一帧的中心角度的变化已被证明在进行定量 MPI 测量时会产生有害的部分扫描伪影。本研究有两个主要目的。(1)证明短扫描 FBP 中存在明显的有害影响,即引入不均匀的空间图像噪声分布;这种不均匀性会导致图像噪声和条纹伪影出乎意料地高,这可能会影响 CT MPI 定量。(2)证明统计图像重建(SIR)算法可以成为解决非均匀空间噪声分布问题的潜在解决方案,并且还可以在 CT MPI 的背景下降低辐射剂量。

方法

本研究使用数值模拟灌注体模和体内动物心肌灌注 CT 扫描的投影数据集。在数值体模中,在每一帧的投影数据中添加了多个泊松噪声的实现,以研究噪声的空间分布。使用 FBP 和 SIR 重建算法对所有数据集的图像进行重建。为了量化噪声的空间分布,在多个感兴趣区域(ROI)中测量均值和标准差,并在时间帧之间进行分析。在体内研究中,使用管电流为 25 和 50 mA 的两个低剂量扫描使用 FBP 和 SIR 进行重建。对两个 ROI 测量了定量灌注指标,即归一化上升斜率(NUS)、心肌血容量(MBV)和第一时刻通过时间(FMT),并与在 500 mA 时进行的高剂量扫描的参考值进行了比较。

结果

使用 FBP 重建的图像显示出高度不均匀的空间噪声分布。这种空间不均匀性导致在时间方向上出现大的波动。在数值体模研究中,在给定的图像中,噪声水平变化了 87%,在远离等中心的 ROI 中,在不同的时间帧之间变化了 110%。不均匀的噪声模式与源轨迹和物体结构有关。相比之下,使用 SIR 重建的图像显示出高度均匀的空间噪声分布,当使用短扫描角度范围时,在时间方向上导致较小的意外噪声波动。在数值体模研究中,在给定的图像中噪声变化小于 37%,在不同的时间帧之间变化小于 20%。此外,SIR 图像中的噪声标准偏差平均为 FBP 图像的一半。在体内研究中,当使用 SIR 而不是 FBP 来重建图像时,低剂量扫描和高剂量扫描之间测量的定量灌注指标之间的偏差得到了缓解。

结论

(1)使用 FBP 重建的图像存在不均匀的空间噪声水平。这种不均匀性是 CT MPI 中短扫描重建造成的有害影响的另一种表现。(2)使用 SIR 重建的图像具有更低且更均匀的噪声水平,因此可以用作解决 FBP 不均匀性的潜在解决方案。(3)鉴于使用 SIR 时灌注度量的准确性提高,可能希望使用统计重建框架来执行低剂量动态 CT MPI。

相似文献

8
Radiation Dose Reduction in CT Myocardial Perfusion Imaging Using SMART-RECON.使用 SMART-RECON 降低 CT 心肌灌注成像中的辐射剂量。
IEEE Trans Med Imaging. 2017 Dec;36(12):2557-2568. doi: 10.1109/TMI.2017.2747521. Epub 2017 Aug 30.

引用本文的文献

6
Myocardial perfusion imaging with cardiac computed tomography: state of the art.心脏 CT 心肌灌注成像:现状。
J Cardiovasc Transl Res. 2013 Oct;6(5):695-707. doi: 10.1007/s12265-013-9499-3. Epub 2013 Aug 21.

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验