顺应性不匹配对动脉移植置换模型中血流紊乱的影响。

Effect of compliance mismatch on flow disturbances in a model of an arterial graft replacement.

作者信息

Miyawaki F, How T V, Annis D

机构信息

Institute of Medical & Dental Bioengineering, University of Liverpool, UK.

出版信息

Med Biol Eng Comput. 1990 Sep;28(5):457-64. doi: 10.1007/BF02441969.

Abstract

Flow disturbances in a model of an interposition graft in an arterial segment were measured using an ultrasound Doppler velocimeter. The effect of the degree of compliance mismatch between a stiff' graft' and compliant 'arterial' segments was investigated. In steady flow, disturbances were detected when the compliance ratio (stiff to compliant segments) was less than or equal to 0.1 and the Reynolds number greater than or equal to 2200. A recirculation zone just downstream of the distal anastomosis was observed at a Reynolds number greater than or equal to 2400. Disturbances were also measured under pulsatile flow which consisted of a time-varying component superimposed on a steady flow component. The time-varying flow component was either quasiphysiological or sinusoidal in shape. The Reynolds number was 500 but the frequency parameter varied from 4.2 to 8.5. Significant disturbances were observed for conduits with compliance ratio less than or equal to 0.19. The disturbance intensity tended to increase as the compliance ratio decreased and the frequency parameter increased. The magnitude of the disturbance was also greater with the quasiphysiological than the sinusoidal input flow waveform.

摘要

使用超声多普勒测速仪测量动脉段间置移植物模型中的血流紊乱情况。研究了硬“移植物”段与顺应性“动脉”段之间顺应性不匹配程度的影响。在稳定流中,当顺应性比(硬段与顺应性段之比)小于或等于0.1且雷诺数大于或等于2200时,检测到紊乱。在雷诺数大于或等于2400时,在远端吻合口下游观察到一个再循环区。还在脉动流下进行了测量,脉动流由叠加在稳定流分量上的时变分量组成。时变流分量的形状要么是准生理的,要么是正弦的。雷诺数为500,但频率参数从4.2变化到8.5。对于顺应性比小于或等于0.19的管道,观察到显著的紊乱。随着顺应性比降低和频率参数增加,紊乱强度趋于增加。与正弦输入流波形相比,准生理输入流波形的紊乱幅度也更大。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验