Melamed Roman, Johnson Kara, Pothen Ben, Sprenkle Mark D, Johnson Pamela Jo
Abbott Northwestern Hospital, Allina Hospitals and Clinics, Minneapolis, Minnesota 55407, USA.
Blood Press Monit. 2012 Oct;17(5):179-83. doi: 10.1097/MBP.0b013e328356e1c7.
Invasive blood pressure monitoring is the most reliable method in critically ill patients. Noninvasive oscillometric pressure monitoring is also widely used in the ICU, and discrepancies between the methods are common. Inclusion of a blood-conserving device may influence the fidelity of the system and reliability of the invasive pressure values. We examined the dynamic response characteristics of the monitoring systems with and without the blood-conserving device, and assessed agreement between invasive and noninvasive blood pressures.
Fifty-three invasive arterial blood pressure monitoring systems were evaluated over an 8-week period. A standard fast flush test was performed. The natural frequency (Fn) and the amplitude ratio of the invasive monitoring systems were calculated. Simultaneous invasive and noninvasive blood pressures were recorded, and the agreement between two methods was evaluated.
The Fn was significantly higher (15.7 ± 5.1 Hz) in the systems without the blood-conserving device than in the systems containing the device (9.1 ± 2.6 Hz). There was no significant difference in the amplitude ratio. Significant differences were observed between invasive and noninvasive pressures, with systolic pressure showing more variability than the mean arterial pressure.
Inclusion of a blood-conserving device in the existing monitoring circuit may affect its dynamic response characteristics. Discrepancies between invasive and noninvasive pressures are common, with systolic pressure showing more variability than the mean arterial pressure. The low Fn of the invasive monitoring system could be a contributing factor to the discrepancy between two methods.
有创血压监测是危重症患者最可靠的方法。无创示波法血压监测在重症监护病房也广泛应用,两种方法之间的差异很常见。加入血液保存装置可能会影响系统的保真度和有创压力值的可靠性。我们研究了有无血液保存装置时监测系统的动态响应特性,并评估了有创和无创血压之间的一致性。
在8周内对53个有创动脉血压监测系统进行评估。进行标准的快速冲洗试验。计算有创监测系统的固有频率(Fn)和振幅比。同时记录有创和无创血压,并评估两种方法之间的一致性。
无血液保存装置的系统中Fn显著更高(15.7±5.1Hz),高于有该装置的系统(9.1±2.6Hz)。振幅比无显著差异。有创和无创血压之间观察到显著差异,收缩压的变异性大于平均动脉压。
在现有的监测回路中加入血液保存装置可能会影响其动态响应特性。有创和无创血压之间的差异很常见,收缩压的变异性大于平均动脉压。有创监测系统的低Fn可能是两种方法之间差异的一个促成因素。