Department of Radiology, C.J. Gorter Center for High Field MRI, Leiden University Medical Center, Leiden, The Netherlands.
Magn Reson Med. 2012 Oct;68(4):1325-31. doi: 10.1002/mrm.24451. Epub 2012 Aug 6.
An annular dielectric resonator made from distilled water has been designed to operate in degenerate quadrature HEM11 modes at 298.1 MHz (7 Tesla). The circularly polarized B1+ field has a high degree of homogeneity throughout a sample placed within the annulus. The sensitivity of the resonator was measured to be essentially identical to that of an eight-rung high-pass birdcage resonator with the same physical dimensions. High resolution in vivo images have been obtained from the human wrist. A new method of electronically detuning the resonator has also been evaluated. The design is extremely simple and rapid to build, with direct applicability to very high field imaging and also potential integration into human and animal hybrid position emission tomography (PET)/MRI and single-photon emission computed tomography (SPECT)/MRI systems due to the lack of conductor attenuation-induced artifacts in the reconstructed nuclear medicine images.
一个由蒸馏水制成的环形介电谐振器被设计用来在 298.1MHz(7 特斯拉)下工作于简并正交 HEM11 模式。在位于环内的样品中,圆极化 B1+场具有高度的均匀性。测量了谐振器的灵敏度,发现其与具有相同物理尺寸的八个梯级高通鸟笼谐振器基本相同。已经从人手腕获得了高分辨率的体内图像。还评估了一种新的电子调谐谐振器的方法。该设计非常简单且快速构建,可直接应用于超高场成像,并且由于在重建的核医学图像中没有导体衰减引起的伪影,因此具有潜在的集成到人和动物混合位置发射断层扫描(PET)/磁共振成像(MRI)和单光子发射计算机断层扫描(SPECT)/ MRI 系统的能力。