Sechopoulos Ioannis, Bliznakova Kristina, Qin Xulei, Fei Baowei, Feng Steve Si Jia
Department of Radiology and Imaging Sciences and Winship Cancer Institute, Emory University School of Medicine, 1701 Upper Gate Drive Northeast, Suite 5018, Atlanta, Georgia 30322, USA.
Med Phys. 2012 Aug;39(8):5050-9. doi: 10.1118/1.4737025.
To compare the estimate of normalized glandular dose in mammography and breast CT imaging obtained using the actual glandular tissue distribution in the breast to that obtained using the homogeneous tissue mixture approximation.
Twenty volumetric images of patient breasts were acquired with a dedicated breast CT prototype system and the voxels in the breast CT images were automatically classified into skin, adipose, and glandular tissue. The breasts in the classified images underwent simulated mechanical compression to mimic the conditions present during mammographic acquisition. The compressed thickness for each breast was set to that achieved during each patient's last screening cranio-caudal (CC) acquisition. The volumetric glandular density of each breast was computed using both the compressed and uncompressed classified images, and additional images were created in which all voxels representing adipose and glandular tissue were replaced by a homogeneous mixture of these two tissues in a proportion corresponding to each breast's volumetric glandular density. All four breast images (compressed and uncompressed; heterogeneous and homogeneous tissue) were input into Monte Carlo simulations to estimate the normalized glandular dose during mammography (compressed breasts) and dedicated breast CT (uncompressed breasts). For the mammography simulations the x-ray spectra used was that used during each patient's last screening CC acquisition. For the breast CT simulations, two x-ray spectra were used, corresponding to the x-ray spectra with the lowest and highest energies currently being used in dedicated breast CT prototype systems under clinical investigation. The resulting normalized glandular dose for the heterogeneous and homogeneous versions of each breast for each modality was compared.
For mammography, the normalized glandular dose based on the homogeneous tissue approximation was, on average, 27% higher than that estimated using the true heterogeneous glandular tissue distribution (Wilcoxon Signed Rank Test p = 0.00046). For dedicated breast CT, the overestimation of normalized glandular dose was, on average, 8% (49 kVp spectrum, p = 0.00045) and 4% (80 kVp spectrum, p = 0.000089). Only two cases in mammography and two cases in dedicated breast CT with a tube voltage of 49 kVp resulted in lower dose estimates for the homogeneous tissue approximation compared to the heterogeneous tissue distribution.
The normalized glandular dose based on the homogeneous tissue mixture approximation results in a significant overestimation of dose to the imaged breast. This overestimation impacts the use of dose estimates in absolute terms, such as for risk estimates, and may impact some comparative studies, such as when modalities or techniques with different x-ray energies are used. The error introduced by the homogeneous tissue mixture approximation in higher energy x-ray modalities, such as dedicated breast CT, although statistically significant, may not be of clinical concern. Further work is required to better characterize this overestimation and potentially develop new metrics or correction factors to better estimate the true glandular dose to breasts undergoing imaging with ionizing radiation.
比较在乳腺钼靶摄影和乳腺CT成像中,使用乳房实际腺体组织分布获得的归一化腺体剂量估计值与使用均匀组织混合近似法获得的估计值。
使用专用乳腺CT原型系统获取20例患者乳房的容积图像,乳腺CT图像中的体素自动分类为皮肤、脂肪和腺体组织。对分类图像中的乳房进行模拟机械压迫,以模拟钼靶摄影采集时的情况。每个乳房的压缩厚度设定为每个患者上次筛查头尾位(CC)采集时达到的厚度。使用压缩和未压缩的分类图像计算每个乳房的容积腺体密度,并创建额外的图像,其中所有代表脂肪和腺体组织的体素被这两种组织的均匀混合物以对应于每个乳房容积腺体密度的比例取代。将所有四张乳房图像(压缩和未压缩;异质和均质组织)输入蒙特卡洛模拟,以估计乳腺钼靶摄影(压缩乳房)和专用乳腺CT(未压缩乳房)期间的归一化腺体剂量。对于乳腺钼靶摄影模拟,使用的X射线光谱是每个患者上次筛查CC采集时使用的光谱。对于乳腺CT模拟,使用了两种X射线光谱,分别对应于目前临床研究中专用乳腺CT原型系统使用的最低和最高能量的X射线光谱。比较每种模态下每个乳房的异质和均质版本的归一化腺体剂量结果。
对于乳腺钼靶摄影,基于均匀组织近似法的归一化腺体剂量平均比使用真实异质腺体组织分布估计的剂量高27%(Wilcoxon符号秩检验p = 0.00046)。对于专用乳腺CT,归一化腺体剂量的高估平均为8%(49 kVp光谱,p = 0.00045)和4%(80 kVp光谱,p = 0.000089)。在乳腺钼靶摄影中只有两例,在管电压为49 kVp的专用乳腺CT中只有两例,与异质组织分布相比,均匀组织近似法导致的剂量估计值较低。
基于均匀组织混合近似法的归一化腺体剂量导致对成像乳房的剂量有显著高估。这种高估影响了绝对剂量估计值的使用,如用于风险估计,并且可能影响一些比较研究,如使用不同X射线能量的模态或技术时。在更高能量的X射线模态(如专用乳腺CT)中,均匀组织混合近似法引入的误差虽然在统计学上显著,但可能在临床上并不值得关注。需要进一步开展工作,以更好地表征这种高估,并可能开发新的指标或校正因子,以更好地估计接受电离辐射成像的乳房的真实腺体剂量。